Thermo Scientific 146i Instruction Manual page 141

Dynamic gas calibrator

Advertisement

Set Defaults
Factory Calibrate Oven
Thermistor
Low and High Points
Thermo Fisher Scientific
The Set Defaults screen is used to reset the configuration values to factory
defaults.
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Gas Therm > Set Defaults.
RESTORE DEFAULT SETTINGS:
FOR PERM OVEN GAS THERM
OPER
MODE
DIAGS ALARM
The Factory Calibrate Oven Thermistor menu is used for calibrating the
permeation oven heater thermistor to either low point, high point, or set
defaults.
In the Main Menu, choose Service > Perm Oven Settings > Factory
Cal Oven Therm.
The Calibrate Oven Thermistor Low Point screen is used to view and set
the permeation oven thermistor resistance. The low and high point screens
function the same way.
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Oven Therm > Low or High Point.
RESTORE
FACTORY CAL OVEN THERM:
>LOW POINT
HIGH POINT
SET DEFAULTS
OPER
MODE
DIAGS ALARM
RESTORE DEFAULT SETTINGS:
FOR PERM OVEN GAS THERM
ARE YOU SURE YOU WANT
PRESS
TO CONFIRM RESTORE
OPER
MODE
DIAGS ALARM
Model 146i Instruction Manual