Removing the
Measurement Bench
and Lowering the
Partition Panel
REMOVE SCREWS
Partition Panel
Loosen Captive Screw
Remove Screw
Thermo Fisher Scientific
The measurement bench can be removed and the partition panel can be
lowered to improve access to connectors and components. Refer to the
following steps when a procedure requires lowering the partition panel (see
Figure
7-3).
Remove Screw
Pass Through
Connectors

Figure 7-3. Removing the Measurement Bench and Lowering the Partition Panel

Equipment Required:
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF and unplug the power cord.
2. Disconnect the plumbing connections at the rear of the measurement
bench.
Removing the Measurement Bench and Lowering the Partition Panel
Remove Screws
SLIDE BENCH OUT
LOWER PANEL
Remove Hidden Screw
Model 146i Instruction Manual
Servicing
7-9