Measurement Interface Board Replacement - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Measurement
Interface Board
Replacement
Thermo Fisher Scientific
Nut driver, 3/16-inch
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the I/O expansion board (optional), if used. See the "I/O
Expansion Board Replacement" procedure in this chapter.
3. Remove the digital output board. See the "Digital Output Board
Replacement" procedure in this chapter.
4. Unplug all connectors from the motherboard. Note connector locations
to facilitate reconnection.
5. Using the nut driver, remove the six standoffs securing the board to the
rear panel.
6. Pop off the motherboard from motherboard support bracket, and
remove the motherboard.
7. To install the motherboard, follow previous steps in reverse.
Use the following procedure to replace the measurement interface board
(Figure
7-10).
Equipment Required:
Measurement interface board
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.

Measurement Interface Board Replacement

Model 146i Instruction Manual
Servicing
7-27