Ambient Temperature Calibration - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Restore Default
Temperature
Calibration
Thermo Fisher Scientific
The Restore Default Calibration screen is used to reset the pressure
calibration configuration values to factory defaults.
Calibration
In the Main Menu, choose Service > Pressure Calibration > Set
Defaults.
RESTORE USER DEFAULTS:
OPER
MODE
DIAGS ALARM
Ambient
The Ambient Temperature Calibration screen is used to view and set the
ambient temperature sensor calibration. The ambient temperature
calibration is visible only when the instrument is in service mode. For more
information on the service mode, see
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Ambient Temp Calibration.
CALIBRATE PRESSURE SPAN:
CURRENTLY:
SET TO:
MOVE CURSOR
CHANGE VALUE
OPER
MODE
DIAGS ALARM
RESTORE
753.0 mmHg
760.0
mmHg ?
SAVE
RESTORE USER DEFAULTS:
ARE YOU SURE YOU WANT
PRESS
TO CONFIRM RESTORE
OPER
MODE
"Service
Mode" earlier in the chapter.
Model 146i Instruction Manual
Operation
Service Menu
RESTORE
TO?x
DIAGS ALARM
3-91