Hardware Configuration; Table B-9. Contrast Levels - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Hardware
Configuration
Thermo Fisher Scientific
This command sets the span index value to index.
Send:
set span index 1
Receive:
set span index 1 ok
temp comp
This command reports whether temperature compensation is on or off.
The example below shows the temperature compensation is off. This
command is only available if the photometer option is installed.
Send:
temp comp
Receive:
temp comp off
set temp comp onoff
These commands turn the temperature compensation on or off. The
example below turns temperature compensation off. This command is only
available if the photometer option is installed.
Send:
set temp comp off
Receive:
set temp comp off ok
contrast
This command reports the screen's level of contrast. The example below
shows the screen contrast is 50%, according to
Send:
contrast
Receive:
contrast 5:50%
set contrast level
This command sets the screen's level of contrast, according to
The example below sets the contrast level to 50%.
Send:
set contrast 5
Receive:
set contrast 5 ok
Table B–9. Contrast Levels
Level
Contrast Level
0
0%
1
10%
2
20%
3
30%
4
40%
5
50%
6
60%
C-Link Protocol Commands

Hardware Configuration

Table
B–9.
Table
B–9.
Model 146i Instruction Manual
B-37