Calibration; Table B-5. Stream Time Values - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Calibration

Thermo Fisher Scientific
set stream time value
This command enables value, where value is to attach or disable time stamp
to streaming data string, according to
attaches a time stamp to streaming data.
Send:
set stream time 0
Receive:
set stream time 0 ok
Table B–5. Stream Time Values
Value
Stream Time
00
Attaches time stamp to streaming data string
01
Disables time stamp to streaming data string
set cal o3 bkg
This command will auto-calibrate the O
shows a successful auto-calibration of the O
only available if the photometer option is installed.
Send:
set cal o3 bkg
Receive:
set cal o3 bkg ok
set cal o3 coef
This command will auto-calibrate O
concentrations. If the mode is incorrect, the instrument responds with
"can't, wrong settings". The example below shows a successful auto-
calibration of the O
coefficient. This command is only available if the
3
photometer option is installed.
Send:
set cal o3 coef
Receive:
set cal o3 coef ok
lamp
This command reports the current photometer lamp setting in percent.
The example below shows that the lamp setting is 10.0%. This command is
only available if the photometer option is installed.
Send:
lamp
Receive:
lamp 10.0 %
set lamp value
This command sets the photometer lamp setting to value, where value is a
floating point number representing a percentage of the photometer lamp
setting. The example below sets the photometer lamp to 10.0%. This
command is only available if the photometer option is installed.
C-Link Protocol Commands
Table
B–5. The example below
background. The example below
3
background. This command is
3
coefficients based on O
3
Model 146i Instruction Manual
Calibration
span gas
3
B-23