Table of Contents IMPORTANT INFORMATION This manual contains important operating and safety information. You must carefully read and understand the contents of this manual prior to the use of this equipment. Safety Information ..............................3 Alert Signals..............................3 Warnings ..............................3 Introduction................................5 Performance Characteristics and Specifications ..................5 Intended Use ..............................5 Function ................................5 Installation ..............................5...
Safety Information Your Thermo Scientific Type 12200 Blood Bank Dri-Bath Alert Signals has been designed with function, reliability, and safety in mind. It is the user’s responsibility to install it in confor- mance with local electrical codes. For safe operation,...
Page 4
AFETY NFORMATION Please note the following WARNINGS: Warning This warning is presented for compliance with California Refer servicing to qualified personnel. Proposition 65 and other regulatory agencies and only applies to the insulation in this product. This product contains refractory ceramic, refractory ceramic fiber or fiberglass insulation, which can produce respirable dust or fibers during disassembly.
37°C — 10 minutes. Designed for continuous service. Intended Use The Type 12200 Blood Bank Dri-Bath provides a con- trolled dry heat environment for 10mm or 12mm test tubes containing in vitro diagnostic specimens or samples for qualitative or quantitative test procedures.
Operation 10mm (-60) Block Models Note Plug the dri-bath into the properly grounded receptacle. Do not place dri-bath in a draft, sun- Insert a calibrated laboratory thermometer* into a 10 x light or near equipment which emits 75mm test tube containing 1.10 mls. of solution. Place the heat or cold as this may result in fluc- test tube in a well.
PERATION 12mm Block Model Note Plug the dri-bath into the properly grounded receptacle. Do not place dri-bath in a draft, sun- Insert a calibrated laboratory thermometer* into a 12 x light or near equipment which emits 75mm test tube containing 1.25 mls. of solution. Place the heat or cold as this may result in fluc- test tube in a well.
Parts List Part No. Description No. Req. 120 Volt Models PC656X1A Control EL656X1A Element BK250X1A Block, Heating 12MM BK250X2A Block, Heating 10 MM (-60 Models) CR598X1 Cord Set PLX97 Light, Pilot...
Ordering Procedures Please refer to the Specification Plate for the complete model number, serial number, and series number when requesting service, replacement parts or in any corre- spondence concerning this unit. All parts listed herein may be ordered from the Thermo Scientific dealer from whom you purchased this unit or can be obtained promptly from the factory.
(ii) misuse (including use inconsistent with written operating instructions for the product), mishandling, contamination, overheating, modification or alteration of the product by any customer or third party or (iii) use of replacement parts that are obtained from a party who is not an authorized dealer of Thermo Scientific prod- ucts.