Operation
Instrument Controls Menu
Data Treatment
Communication
3-48
Model 146i Instruction Manual
The Data Treatment screen is used to select the data type for the selected
record: whether the data should be averaged over the interval, the
minimum or maximum used, or the current value logged. Data treatment
doesn't apply to all data, just to the concentration measurement. All other
data points log the current value at the end of the interval.
In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging > Data Treatment.
SET LREC DATA TYPE:
The Communication Settings menu is used with communications control
and configuration.
Settings
In the Main Menu, choose Instrument Controls > Communication
Settings.

COMMUNICATION SETTINGS:

>BAUD RATE
Baud Rate
The Baud Rate screen is used to set the RS-232/RS-485 interface baud rate.
Baud rates of 1200, 2400, 4800, and 9600, 19200, 38400, 57600, and
115200 are available.
In the Main Menu, choose Instrument Controls > Communication
Settings > Baud Rate.
CURRENTLY:
AVG
SET TO:
MAX
CHANGE VALUE
OPER
MODE
DIAGS ALARM
INSTRUMENT ID
COMMUNICATION PROTOCOL
STREAMING DATA CONFIG
RS-232/RS-485 SELECTION
TCP/IP SETTINGS
OPER
MODE
DIAGS ALARM
?
SAVE