C-Link Protocol Commands
Measurement Configuration
B-32
Model 146i Instruction Manual
This command sets the tank concentration for gas gas. Here conc is stored
in ppb.
Note Changing this value may affect allowable limits for gas span
concentration and gas span flow. It is advisable to check these settings.
Send:
set gas a tank conc 123.45
Receive:
set gas a tank conc 123.45 ok
l1
l2
l3
l4
l5
These commands report the percentage of the ozonator lamp drive for each
of the five custom levels. The example below shows that custom level 1 is
10%. This command is only available if the ozonator option is installed.
Also, if the photometer option is installed, it must be configured for
internal or external measurement mode.
Send:
l1
Receive:
l1 10%
set l1 value
set l2 value
set l3 value
set l4 value
set l5 value
These commands set the percentage of the ozonator lamp drive for each of
the five custom levels. The example below sets the custom level 1 to 10%.
This command is only available if the ozonator option is installed. Also, if
the photometer option is installed, it must be configured for internal or
external measurement mode.
Send:
set l1 10
Receive:
set l1 10 ok
l1 conc
l2 conc
l3 conc
l4 conc
l5 conc
These commands report the concentration setting of each of the five
custom levels. The example below shows that custom level 1 is 500 ppb.
This command is only available if the ozonator and photometer options are
installed and the photometer is configured for ozonator control mode.