Download Print this page

Ozonator Board Replacement (Ozonator Option); Figure 7-13. Replacing The Ozonator - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Ozonator Board
Replacement
(Ozonator option)
Thermo Fisher Scientific
7. Re-install the instrument cover.
Capillary

Figure 7-13. Replacing the Ozonator

Use the following procedure to replace the ozonator board
Equipment Required:
Ozonator board
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the connection cables from the board.
3. Remove the two screws from the top of the board.
4. Pop off the board from the bottom mounting studs and remove the
board.
5. To install the new ozonator board, follow previous steps in reverse.

Ozonator Board Replacement (Ozonator option)

(Figure
Model 146i Instruction Manual
Servicing
7-10).
7-45

Advertisement

loading