Download Print this page

Thermo Scientific 146i Instruction Manual page 316

Dynamic gas calibrator

Advertisement

C-Link Protocol Commands
Hardware Configuration
B-38
Model 146i Instruction Manual
Level
Contrast Level
7
70%
8
80%
9
90%
10
100%
date
This command reports the current date. The example below reports the
date as December 1, 2004.
Send:
date
Receive:
date 12-01-04
set date mm-dd-yy
mm = month
dd = day
yy = year
This command sets the date of the analyzer's internal clock. The example
below sets the date to December 1, 2004.
Send:
set date 12-01-04
Receive:
set date 12-01-04 ok
set default params
This command sets all the parameters to their default values. This does not
affect the factory-calibrated parameters.
Send:
set default params
Receive:
set default params ok
save
set save params
This command stores all current parameters in FLASH memory. It is
important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure. The example below saves the parameters to FLASH
memory.
Send:
set save params
Receive:
set save params ok
time
This command reports the current time (24-hour time). The example
below reports that the internal time is 2:15:30 pm.
Thermo Fisher Scientific

Advertisement