C-Link Protocol Commands

Alarms

B-10
Model 146i Instruction Manual
Figure B–1. Flags
alarm bench lamp temp min
Alarms
alarm bench lamp temp max
These commands report the photometer bench lamp temperature alarm
minimum and maximum value current settings. The example below reports
the bench lamp temperature alarm minimum value is 52 °C. These
commands are only available if the photometer option is installed.
Send:
Receive:
set alarm bench lamp temp min value
set alarm bench lamp temp max value
These commands set the photometer bench lamp temperature minimum
and maximum values to value, where value is a floating-point number
representing bench lamp temperature alarm limits in degrees C. The
example below sets the bench lamp temperature alarm maximum value to
58 °C. These commands are only available if the photometer option is
installed.
Send:
Receive:
alarm bench lamp temp min
alarm bench lamp temp min 52.0 deg C
set alarm bench lamp temp max 58
set alarm bench lamp temp max 58 ok
Thermo Fisher Scientific