Measurement Configuration; Table B-6. Gas Index Values - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

C-Link Protocol Commands

Measurement Configuration

Measurement
Configuration
B-28
Model 146i Instruction Manual
Send:
screen
Receive:
screen
This is an I series
Instrument. Screen
Information not
available
gas index
This command reports the current gas index. This selects the gas that will
be diluted, according to
is currently selected.
Send:
gas index
Receive:
gas index 1
set gas index index
index = 0-6
This command sets the current gas index value to index. The example
below selects Gas D, according to
Send:
set gas index 4
Receive:
set gas index 4 ok
Table B–6. Gas Index Values
Value
Gas Index
0
None
1
Gas A
2
Gas B
3
Gas C
4
Gas D
5
Gas E
6
Gas F
sel gas gas man
gas = | a | b | c | d | e | f |
This command selects the gas at the level preset for manual.
Send:
sel gas a man
Receive:
sel gas a man ok