Analog Output Calibrate
Analog Output Calibrate
Thermo Fisher Scientific
In the Main Menu, choose Service > Analog Out Calibration.
The Analog Output Calibrate Zero screen is used to calibrate the zero state
Zero
of the selected analog output. The operator must connect a meter to the
output and adjust the output until it reads 0.0 V on the meter.
In the Main Menu, choose Service > Analog Out Calibration > Select
Channel > Calibrate Zero.
The Analog Output Calibrate Full-Scale screen is used to calibrate the full-
Full-Scale
scale state of the selected analog output. The operator must connect a
meter to the output and adjust output until it reads the value shown in the
set output to: field.
In the Main Menu, choose Service > Analog Out Calibration > Select
Channel > Calibrate Full Scale.
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
VOLTAGE CHANNEL 6
CURRENT CHANNEL 1
OPER
MODE
DIAGS ALARM
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
OPER
MODE
DIAGS ALARM
ANALOG OUTPUT CAL:
CONNECT METER TO OUTPUT!
SELECTED OUTPUT:
SET TO:
SAVE VALUE
SET OUTPUT TO:
OPER
MODE
DIAGS ALARM
ZERO
V1
100
INC/DEC
0.0 V
Model 146i Instruction Manual
Operation
Service Menu
3-93