Thermo Scientific 146i Instruction Manual page 293

Dynamic gas calibrator

Advertisement

Thermo Fisher Scientific
intern temp
This command reports the current internal instrument temperature, in
degrees C. The example below shows that temperature compensation is on
and that the internal temperature is 27.2 °C.
Send:
intern temp
Receive:
intern temp 27.2 deg C
lamp temp
This command reports the photometer bench lamp temperature. The
example below reports the current bench lamp temperature is 55.2 °C. This
command is only available if the photometer option is installed.
Send:
lamp temp
Receive:
lamp temp 55.2 deg C
lamp voltage bench
This command reports the photometer bench lamp voltage in volts. This
command is only available if the photometer option is installed.
Send:
lamp voltage bench
Receive:
lamp voltage bench 10.3 V
lamp voltage o3
This command reports the ozonator lamp voltage in volts. This command
is only available if the photometer option is installed.
Send:
lamp voltage o3
Receive:
lamp voltage o3 12.3 V
oz lamp temp
This command reports the current ozonator lamp temperature. The
example below shows the current ozonator temperature is 68.7 °C. This
command is only available if the photometer option is installed.
Send:
oz lamp temp
Receive:
oz lamp temp 68.7 deg C
photometer temp
This command reports the current photometer bench temperature. The
first reading is the temperature being used in instrument calculations. The
second temperature is the actual temperature being measured. If
temperature compensation is on, then both temperature readings are the
same. If temperature compensation is off, a temperature of 0.0 °C is shown
as the default temperature even though the actual internal temperature is
31.2 °C The example below shows that temperature compensation is on
and that the internal temperature is 31.2 °C.
C-Link Protocol Commands
Diagnostics
Model 146i Instruction Manual
B-15