Keys/Display

Thermo Fisher Scientific
resp coef
This command reports the current ozonator response coefficient when
under photometer control. The example below reports that the ozonator
response coefficient is 1.000. This command is only available if the
photometer option is installed.
Send:
resp coef
Receive:
resp coef 1.000
set resp coef value
This command sets the ozonator response coefficient to user-defined values
to value, where value is a floating-point representation of the coefficient.
The example below sets the ozonator response coefficient to 1.500. This
command is only available if the photometer option is installed.
Send:
set resp coef 1.500
Receive:
set resp coef 1.500 ok
sp conc
This command reports photometer span concentration. If the mode is
incorrect, the instrument responds with "can't, wrong settings". The
example below reports the span gas concentration of 50 ppb. This
command is only available if the photometer option is installed.
Send:
sp conc
Receive:
sp conc 5.000E+01 ppb
set sp conc value
This command sets the span concentrations to user-defined values to value,
where value is a floating-point representation of the span concentration in
ppb. The example below sets the span concentration to 50 ppb. This
command is only available if the photometer option is installed.
Send:
set sp conc 50
Receive:
set sp conc 50 ok
push button
do
down
en
enter
he
help
le
left
me
C-Link Protocol Commands
Keys/Display
Model 146i Instruction Manual
B-25