Download Print this page

Program Cycle; Period Hours - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Program Cycle

Period Hours

Thermo Fisher Scientific
PROGRAM:
>PROGRAM CYCLE
PERIOD HOURS
NEXT CYCLE 19JAN06 08:27
NEXT EVENT
EVENTS
OPER
MODE
The Program Cycle screen is used turn the program mode on or off. When
enabled, the status bar shows the current status of programmed events,
either ACTIVE or INACTIVE. Inactive means enabled but not currently
running, and active means running current event. When disabled, the
program will not be run at its assigned interval. This screen is also used to
cancel a running program.
In the Main Menu, choose > Program > Program Cycle.
PROGRAM CYCLE STATUS
CURRENTLY:
SET TO:
OPER
MODE
The Period Hours screen is used select how many hours starting between
cycles of the program. Valid values are 0 to 999. If 0 is selected, the
program will run only once then it will be disabled.
In the Main Menu, choose > Program > Period Hours.
PROGRAM CYCLE PERIOD:
CURRENTLY:
SET TO:
CHANGE VALUE
OPER
MODE
ENABLED
24
#31 15:00
DIAGS ALARM
DISABLED
ENABLED?
TOGGLE VALUE
DIAGS ALARM
024 HRS
120 HRS?
MOVE CURSOR
SAVE
DIAGS ALARM
Model 146i Instruction Manual
Operation
Program Menu
3-37

Advertisement

loading