Dynamic gas calibrator

Advertisement

Thermo Fisher Scientific
alarm bench temp min
alarm bench temp max
These commands report the photometer bench temperature alarm
minimum and maximum value current settings. The example below reports
the bench temperature alarm minimum value is 15.0 °C. These commands
are only available if the photometer option is installed.
Send:
alarm bench temp min
Receive:
alarm bench temp min 15.0
set alarm bench temp min value
set alarm bench temp max value
These commands set the photometer bench temperature minimum and
maximum values to value, where value is a floating-point number
representing bias voltage alarm limits in degrees C. The example below sets
the bench temperature alarm maximum value to 35 °C. These commands
are only available if the photometer option is installed.
Send:
set alarm bench temp max 35
Receive:
set alarm bench temp max 35 ok
alarm conc o3 min
alarm conc o3 max
These commands report the photometer O
and maximum value current setting. The example below reports the O
concentration alarm minimum is .20 ppb. These commands are only
available if the photometer option is installed.
Send:
alarm conc o3 min
Receive:
alarm conc o3 min 20.0 ppm
set alarm conc o3 min value
set alarm conc o3 max value
These commands set the photometer O
maximum values to value, where value is a floating-point representation of
the concentration alarm limits in ppb. The example below sets the O
concentration alarm maximum value to 5. These commands are only
available if the photometer option is installed.
Send:
set alarm conc o3 max 5
Receive:
set alarm conc o3 max 5 ok
alarm intensity min
alarm intensity max
These commands report the photometer detector intensity alarm minimum
and maximum value current settings. The example below reports the
intensity alarm minimum value is 45000 Hz. These commands are only
available if the photometer option is installed.
C-Link Protocol Commands
concentration alarm minimum
3
concentration minimum and
3
Model 146i Instruction Manual
Alarms
3
3
B-11