Thermo Scientific 146i Instruction Manual page 139

Dynamic gas calibrator

Advertisement

Calibrate Oven Thermistor
Permeation Oven Selection
Thermo Fisher Scientific
Resistor
The Calibrate Gas Thermistor Resistor screen is used to view and set the
permeation oven gas thermistor resistance to a known resistor value.
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
Thermistor > Known Resistor.
CAL GAS THERM (RESISTOR):
CURRENTLY:
OPER
The Calibrate Oven Thermistor screen is used to view and set the
permeation oven thermistor resistance to a known resistor value.
In the Main Menu, choose Service > Perm Oven Settings > Cal Oven
Thermistor.
CAL OVEN THERM (RESISTOR):
CURRENTLY:
OPER
The Permeation Oven Setpoint screen is used to select the permeation oven
state to not present, or the desired temperature choice of 30, 35, and 45 °C.
In the Main Menu, choose Service > Perm Oven Settings > Perm Oven
Selection.
3850 Ohms
SET TO:
04000
MOVE CURSOR
CHANGE VALUE
MODE
DIAGS ALARM
3850 Ohms
SET TO:
04000
MOVE CURSOR
CHANGE VALUE
MODE
DIAGS ALARM
Ohms
SAVE
Ohms
SAVE
Model 146i Instruction Manual
Operation
Service Menu
3-97