Calibrate Oven Thermistor
Permeation Oven Selection
Thermo Fisher Scientific
Resistor
The Calibrate Gas Thermistor Resistor screen is used to view and set the
permeation oven gas thermistor resistance to a known resistor value.
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
●
Thermistor > Known Resistor.
CAL GAS THERM (RESISTOR):
CURRENTLY:
OPER
The Calibrate Oven Thermistor screen is used to view and set the
permeation oven thermistor resistance to a known resistor value.
In the Main Menu, choose Service > Perm Oven Settings > Cal Oven
●
Thermistor.
CAL OVEN THERM (RESISTOR):
CURRENTLY:
OPER
The Permeation Oven Setpoint screen is used to select the permeation oven
state to not present, or the desired temperature choice of 30, 35, and 45 °C.
In the Main Menu, choose Service > Perm Oven Settings > Perm Oven
●
Selection.
3850 Ohms
SET TO:
04000
MOVE CURSOR
CHANGE VALUE
MODE
DIAGS ALARM
3850 Ohms
SET TO:
04000
MOVE CURSOR
CHANGE VALUE
MODE
DIAGS ALARM
Ohms
SAVE
Ohms
SAVE
Model 146i Instruction Manual
Operation
Service Menu
3-97