Datalogging - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

C-Link Protocol Commands

Datalogging

Datalogging
B-16
Model 146i Instruction Manual
Send:
photometer temp
Receive:
photometer temp 31.2 deg C, actual 31.2
program
This command reports programmed events status.
Send:
program
Receive:
program on
set program onoff
This command sets programmed events on or off. If set to off, programmed
events will not be triggered at their normal interval and if programmed
events are currently running, they will be aborted. All the other set
commands will be disabled if the programmed events are running, except
for this command. If set to on, programmed events will be triggered at their
normal interval.
Send:
set program on
Receive:
set program on ok
clr records
This command will clear all long and short records that have been saved.
Send:
clear records
Receive:
clear records ok
set clr lrecs
set clr srecs
These commands will clear only the long records or only the short records
that have been saved. The example below clears short records.
Send:
set clr srecs
Receive:
set clr srecs ok
list din
list dout
These commands report the current selection for the digital outputs in the
format. Output no Index number variable name active state. The active
state for digital outputs is open or closed. The active state for digital inputs
is high or low.
Send:
list dout
Receive:
list dout
output index variable state
1 1 GAS BIT 1 open
2 2 GAS BIT 2 open
3 3 GAS BIT 3 open
4 4 SPAN BIT 1 open