Pressure Transducer Replacement (Photometer Option) - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Pressure Transducer
Replacement
(Photometer option)
Thermo Fisher Scientific
Philips screwdriver
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Unplug the solenoid electrical cable.
3. Disconnect all lines from defective solenoid.
4. Remove the solenoid and replace with the new solenoid.
5. Reconnect the electrical connection and the lines.
6. Check for leaks after replacing any solenoid. Refer to "Leak Test and
Pump Checkout" in this chapter.
Use the following procedure to replace the pressure transducer. Refer to
"Component Layout"
the instrument, and refer to
need the pressure transducer part number.
Equipment Required:
Pressure transducer assembly
Philips screwdriver, short
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect plumbing from the pressure transducer assembly. Note the
plumbing connections to facilitate reconnection.
3. Disconnect the pressure transducer PRESS connector from the
measurement interface board.

Pressure Transducer Replacement (Photometer option)

on
page 7-8
to locate the pressure transducer within
"Replacement Parts List"
Servicing
on
page 7-5
if you
Model 146i Instruction Manual
7-39