Bench Temperature Calibration; Analog Output Calibration - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Operation
Service Menu
Bench Temperature
Calibration
Analog Output
Calibration
3-92
Model 146i Instruction Manual
CALIBRATE AMBIENT TEMP:
CURRENTLY:
SET TO:
CHANGE VALUE
OPER
MODE
The Bench Temperature Calibration screen is used to view and set the
bench temperature sensor calibration. This temperature is used for
compensation of the photometer option's ozonator measurement. The
temperature calibration is visible only if the photometer option is installed
and when the instrument is in service mode. For more information on the
service mode, see
"Service
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Bench Temp Calibration.
CALIBRATE BENCH TEMP:
CURRENTLY:
SET TO:
CHANGE VALUE
OPER
MODE
The Analog Output Calibration menu is a selection of 6 voltage channels
and 6 current channels (if I/O expansion board option is installed) to
calibrate, and allows the user to select the calibration action zero or span.
The analog output calibration is visible only when the instrument is in
service mode. For more information on the service mode, see
Mode" earlier in the chapter.
Note This adjustment should only be performed by an instrument service
technician.