Download Print this page

Thermo Scientific 146i Instruction Manual page 35

Dynamic gas calibrator

Advertisement

Thermo Fisher Scientific
Pins 3, 4, and 23 select a gas. Pins 2, 5, 10, 13, 16, 18, 19, 22, 26, 32, 35,
and 37 are ground.
Note Gases D, E, and F are available only when the instrument is equipped
with the 6-gas option.
Gas Bit 3
Gas Bit 2
Pin 23
Pin 4
0
0
0
0
0
1
0
1
1
0
1
0
1
1
Pins 24, 6, and 25 select a gas span level gas.
Span Bit 3
Span Bit 2
Pin 25
Pin 6
0
0
0
0
0
1
0
1
1
0
1
0
1
1
Pins 7, 8, 27, and 9 select an ozonator or permeation oven level gas.
Oz/Perm Bit 4
Oz/Perm Bit 3
Pin 9
Pin 27
0
0
0
0
0
0
0
0
0
1
0
1
Gas Bit 1
Gas Type
Pin 3
0
Gas Off
1
Gas A
0
Gas B
1
Gas C
0
Gas D
1
Gas E
0
Gas F
Span Bit 1
Span Level
Pin 24
0
Span 0
1
Span 1
0
Span 2
1
Span 3
0
Span 4
1
Span 5
0
Manual
Oz/Perm 2
Oz/Perm Bit 1
Pin 8
Pin 7
0
0
0
1
1
0
1
1
0
0
0
1
Model 146i Instruction Manual
Installation
Remote Interconnections
Ozonator or
Permeation
Oven Levell
Ozon/Perm Off
Ozon Manual
Ozon 1
Ozon 2
Ozon 3
Ozon 4
2-7

Advertisement

loading