Pressure Regulator Capillary Replacement; Figure 5-2. Replacing The Ozonator Capillary - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Pressure Regulator
Capillary
Replacement
Thermo Fisher Scientific
3. Insert the new capillary and reconnect the fittings.
Capillary

Figure 5-2. Replacing the Ozonator Capillary

Use the following procedure to replace the pressure regulator capillary
(Figure
5-3).
Equipment Required:
Pressure regulator capillary
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Unscrew the knurled fitting, separate the components, and remove the
capillary
(Figure
5-3).
Preventive Maintenance

Pressure Regulator Capillary Replacement

Model 146i Instruction Manual
5-7