Preventive Maintenance

Pump Capillary Replacement (Photometer option)

Pump Capillary
Replacement
(Photometer option)
5-8
Model 146i Instruction Manual
3. Insert the new capillary and reconnect the fittings.

Figure 5-3. Replacing the Pressure Regulator Capillary

For instruments with the photometer option, use the following procedure to
replace the pump capillary
Equipment Required:
Pump capillary
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Loosen the fitting connecting the capillary to the pump and remove the
capillary.
(Figure
5-4).
Capillary
(Figure
5-4).