Reaction Chamber/Mixing Chamber Replacement; Figure 7-15. Replacing The Glass Chamber Assembly - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Servicing

Reaction Chamber/Mixing Chamber Replacement

Reaction
Chamber/Mixing
Chamber
Replacement
7-48
Model 146i Instruction Manual
Permeation
Oven Top
Glass
Chamber
Assembly
Permeation
Oven Bottom

Figure 7-15. Replacing the Glass Chamber Assembly

Use the following procedure to replace the reaction chamber or mixing
chamber.
Equipment Required:
Reaction chamber or mixing chamber
Allen wrench, 7/64-inch
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the plumbing.
3. Remove two (2) Allen screws from the top of the holder assembly.
4. Remove two (2) Allen screws from the side of the holder assembly.
See Procedure
Oven Top
Chamber Top
See Expanded View
Expanded View