I/O Configuration; Table B-11. Analog Current Output Range Values - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

I/O Configuration

Thermo Fisher Scientific
Send:
program no
Receive:
program no iSeries 146i 01.00.00.100
set layout ack
This command disables the stale layout/layout change indicator ('*') that is
attached to each response if the layout has changed.
Send:
set layout ack
Receive:
set layout ack ok
analog iout range channel
This command reports the analog current output range setting for channels,
where channel must be between 1 and 6, inclusive. The example below
reports current output channel 4 to the 4-20 mA range, according to
B–11. This command responds with "feature not enabled" if the I/O
expansion board is not detected.
Send:
analog iout range 4
Receive:
analog iout range 4 2
set analog iout range channel range
This command sets analog current output channel to the channel range
where channel is between 1 and 6 inclusive, and range is set according to
Table
B–11. The example below sets current output channel 4 to the 0-20
mA range. This command responds with "feature not enabled" if the I/O
expansion board is not detected.
Send:
set analog iout range 4 1
Receive:
set analog iout range 4 1 ok
Table B–11. Analog Current Output Range Values
Range
1
2
0 [cannot be set to this, but may report]
analog vin channel
This command retrieves the analog voltage input channel data, both the
calculated value and the actual voltage. In the example below, the
"calculated" value of channel 1 is 75.325 degrees F, volts are 2.796. This
command responds with "feature not enabled" if the I/O expansion board
is not detected.
C-Link Protocol Commands
I/O Configuration
Output Range
0-20 mA
4-20 mA
Undefined
Model 146i Instruction Manual
Table
B-43