Download Print this page

Thermo Scientific 146i Instruction Manual page 309

Dynamic gas calibrator

Advertisement

Thermo Fisher Scientific
set gas gas span conc spanindex value
gas = | a | b | c | d | e | f |
spanindex= span index 0-6 (1-5 reports the span concentration for each
span level and 6 reports the span concentration for manual mode)
value = float
This command sets the target span concentration value for the specified gas
gas and the span level spanindex. The concentration value is in ppb.
Note Changing span concentration may affect allowable limits for gas span
flow. It is advisable to check this setting.
Send:
set gas a span conc 1 123.45
Receive:
set gas a span conc 1 123.45 ok
gas gas span flow
gas = | a | b | c | d | e | f |
spanindex = span index 0-6 (0 is for zero flow)
This command reports the target total flow for the specified gas gas and
span level spanindex.
Send:
gas a span flow 1
Receive:
gas a span flow 1 123.45
set gas gas span flow value
gas = | a | b | c | d | e | f |
spanindex = span index 0-6 (0 is for zero flow)
value = float
This command sets the target total flow value for the specified gas gas and
span level spanindex.
Send:
set gas a span flow 1 123.45
Receive:
set gas a span flow 1 123.45 ok
gas gas tank conc
gas = | a | b | c | d | e | f |
This command reports the tank concentration for gas gas. Here
concentration is reported in ppb.
Send:
gas a tank conc
Receive:
gas a tank conc 123.45
set gas gas tank conc conc
gas = | a | b | c | d | e | f |
conc = float