Dynamic gas calibrator

Advertisement

Thermo Fisher Scientific
gas name gas
gas = | a | b | c | d | e | f |
This command reports the gas name for gas gas.
Send:
gas a name
Receive:
gas a name abcd
set gas name gas name
gas = | a | b | c | d | e | f |
name = 5 character string
This command sets the gas name for gas gas to a name. 1-5 characters are
allowed.
Send:
set gas a name abcd
Receive:
set gas a name abcd ok
sel gas off
This command turns off all gas solenoids.
Send:
sel gas off
Receive:
sel gas off ok
gas gas params
gas = | a | b | c | d | e | f |
This command returns the parameters for the specified gas.
Send:
gas a params
Receive:
gas a params
tank ppm 10.000
span 0 zflow 1800.000
span 1 tflow 1000.000 ppm 0.400
span 2 tflow 1000.000 ppm 0.800
span 3 tflow 1000.000 ppm 1.200
span 4 tflow 1000.000 ppm 1.600
span 5 tflow 1000.000 ppm 0.400
manual zflow 1000.000 gflow 30.50
gas gas solenoid
gas = | a | b | c | d | e | f |
This command reports the solenoid number, according to
which is used when the gas gas is selected. The example below shows that
solenoid A is associated with gas A.
Send:
gas a solenoid
Receive:
gas a solenoid 1
C-Link Protocol Commands
Measurement Configuration
Table
B-7,
Model 146i Instruction Manual