Download Print this page

Thermo Scientific 146i Instruction Manual page 296

Dynamic gas calibrator

Advertisement

C-Link Protocol Commands
Datalogging
B-18
Model 146i Instruction Manual
erec
This command returns a brief description of the main operating conditions
at the time the command is issued (i.e. dynamic data). The example below
shows a typical response. The format is defined by the current settings of
"format" and "erec format" commands. For details on how to decode the
flag fields within these records, see the "flags" command.
Send:
erec
Receive:
erec
14:08 12-15-05 C000000 0.000 4.999.900 999.900 999.900
999.900 50.000 0.000 0.000 0.000 0.000 1.000 90 6.000 1
1 0.000 1 0 0.000 0 0 0.000
lrec
srec
lrec xxxx yy
srec xxxx yy
lrec aa:bb oo-pp-qq yy
srec aa:bb oo-pp-qq yy
xxxx = the number of past records
yy = the number of records to return (1 to 10)
aa = hours (01 to 24)
bb = minutes (01 to 59)
oo = month (01 to 12)
pp = day (01 to 31)
qq = year
These commands output long or short records. The output format is
determined by the "set lrec format" and "set srec format" commands. The
logging time is determined by the "set lrec per" and "set srec per"
commands.
In the following example, there are 740 long records currently stored in
memory. When the command lrec 100 5 is sent, the instrument counts
back 100 records from the last record collected (record 740), and then
returns 5 records: 640, 641, 642, 643, and 644. For details on how to
decode the flag fields within these records, see the "flags" command.
Send:
lrec 5
Receive:
lrec 100 5
10:00 12-15-05 flags C000000 conc 0.000 tgflo 30.000
agflo 0.000 tzflo 1000.000 azflo 0.000 ttflo 1030.000
atflo 0.000 pres 0.000
11:00 12-15-05 flags C000000 conc 0.000 tgflo 30.000
agflo 0.000 tzflo 1000.000 azflo 0.000 ttflo 1030.000
atflo 0.000 pres 0.000
12:00 12-15-05 flags C000000 conc 0.000 tgflo 30.000
agflo 0.000 tzflo 1000.000 azflo 0.000 ttflo 1030.000
atflo 0.000 pres 0.000