Intensity Check

Reference Gas
Sample Gas
Thermo Fisher Scientific
The Intensity Check menu is used to manually control the flow of
reference of sample gas through the photometer. This enables the intensity
and noise reading of the detector to be determined with either reference or
sample gas flow. Selecting any of these menu choices will disturb the analog
outputs. The intensity check is visible only if the photometer option is
installed and when the calibrator is in service mode. For more information
on the service mode, see
In the Main Menu, choose Service > Intensity Check.
INTENSITY CHECK:
>REFERENCE GAS
SAMPLE GAS
OPER
MODE
The Reference Gas screen (read only) sets the valves to route the reference
gas through the photometer and displays the intensity and noise.
In the Main Menu, choose Service > Intensity Check > Reference Gas.
PHOTOMETER REFERENCE GAS:
INTENSITY:
NOISE:
OPER
MODE
The Sample Gas screen (read only) sets the valves to route the sample gas
through the photometer and displays the intensity and noise.
In the Main Menu, choose Service > Intensity Check > Sample Gas.
"Service
Mode" earlier in the chapter.
DIAGS ALARM
98425 Hz
1.4
DIAGS ALARM
Operation
Service Menu
Model 146i Instruction Manual
3-101