Thermo Scientific 146i Instruction Manual page 104

Dynamic gas calibrator

Advertisement

Operation
Instrument Controls Menu
Analog Output
Configuration
Select Output Range
3-62
Model 146i Instruction Manual
The Analog Output Configuration menu displays a list of the analog
output channels available for configuration. Channel choices include all
voltage channels, all current channels, voltage channels 1-6, and current
channels 1-6 (if the I/O expansion board option is installed). Configuration
choices include selecting range, setting minimum/maximum values, and
choosing signal to output.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config.
OUTPUT CHANNELS:
>ALL VOLTAGE CHANNELS
ALL CURRENT CHANNELS
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
OPER
MODE
ANALOG OUTPUT CONFIG:
>SELECT RANGE
SET MINIMUM VALUE
SET MAXIMUM VALUE
CHOOSE SIGNAL TO OUTPUT
OPER
MODE
The Select Output Range screen is used to select the hardware range for the
selected analog output channel. Possible ranges for the voltage outputs are:
0-100 mV, 0-1, 0-5, 0-10 V.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > Select Channel > Select Range.
DIAGS ALARM
DIAGS ALARM
Thermo Fisher Scientific