Table B-13. Default Output Assignment - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

Thermo Fisher Scientific
This command assigns digital input channel (1-16) to activate the action
indicated by index (1-35), when the input transitions to the designated state
(high or low). Use "list din var" command to obtain the list of supported
index values and corresponding actions.
Send:
set din 5 9 high
Receive:
set din 1 9 high ok
dout channel
This command reports the index number and output variable and the
active state assigned to output channel. The example below reports the
input 2 to be assigned an index number 2 corresponding to "local/remote"
with the active state being open.
Send:
dout 2
Receive:
dout 2 2 LOCAL/REMOTE open
set dout channel index state
This command assigns digital output channel to be assigned to the action
associated with index, and assigns it an active state of state (open or closed).
Send:
set dout 2 2 open
Receive:
set dout 2 2 open ok
dtoa channel
This reports the outputs of the 6 or 12 Digital to Analog converters,
according to
Table
B–13. The example below shows that the D/A #1 is
97.7% full-scale.
Send:
dtoa 1
Receive:
dtoa 1 97.7%
Note All channel ranges are user definable. If any customization has been
made to the analog output configuration, the default selections may not
apply.
Table B–13. Default Output Assignment
D to A
Function
1
Voltage Output
2
Voltage Output
3
Voltage Output
4
Voltage Output
5
Voltage Output
6
Voltage Output
Single Range
Dual Range
O
Not Used
3
O
Not Used
3
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Model 146i Instruction Manual
C-Link Protocol Commands
I/O Configuration
Auto Range
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
B-45