Operation
Program Menu

Next Cycle

3-38
Model 146i Instruction Manual
The Next Cycle screen is used view and change the start date and time of
the program in 24-hour time.
In the Main Menu, choose > Program > Next Cycle.
NEXT CYCLE START TIME:
18 JAN 2006
PRESS
TO EDIT
OPER
MODE
DIAGS ALARM

Events

The Events menu lets you configure events that are part of the program
cycle. The number of events and total time of enabled events are displayed.
The target values on the right side of the display, is an abbreviated line that
describes the settings for that specific event, and should show the desired
values based on the displayed operation parameters on the left side. Specific
parameters include: gas (A-F, Zero), span level for that gas (1-5, Zero),
ozonator and permeation levels (1-5, OFF), photometer mode (I=internal,
E=external, C=control), and event duration time in minutes. Selecting an
event allows detailed setup of that event. A maximum of 32 events can be
controlled from this menu.
Note When events are running, none of the settings may be changed via
the front panel interface, except for the program cycle enable/disable
screen.
In the Main Menu, choose > Program > Events.
12:07
PROGRAM CYCLE EVENTS:
TOT EVENTS
>EVENT 1
GA2 03 P1 I
EVENT 2
GC4 03
EVENT 3
GZ
EVENT 4
EVENT 5
EVENT 6
OPER
MODE
DIAGS ALARM
NEXT CYCLE START TIME:
18 JAN 2006
SETTING: DAYS
SET MONTHS
OPER
MODE
2 TIME 0:06
1
C
5
P4
OFF
OFF
OFF
OFF
12:07
?
CHANGE VALUE
SAVE VALUE
DIAGS ALARM
Thermo Fisher Scientific