Thermo Scientific 146i Instruction Manual page 101

Dynamic gas calibrator

Advertisement

Instrument Action
External Solenoid
Thermo Fisher Scientific
DIGITAL INPUT SETUP:
>LOGIC STATE
INSTRUMENT STATE
OPER
The Instrument Action screen allows the user to choose the instrument
action that is tied to the selected digital input.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Settings > Select Relay > Instrument Action.
CHOOSE ACTION:
>NONE
GAS BIT 1
GAS BIT 2
GAS BIT 3
SPAN BIT 1
SPAN BIT 2
SPAN BIT 3
OPER
O3/PERM BIT 1
O3/PERM BIT 2
O3/PERM BIT 3
O3/PERM BIT 4
PHOTO BIT 1
PHOTO BIT 2
AOUTS TO ZERO
AOUTS TO FS
The Model 146i can drive up to eight external solenoids. You can select the
external solenoids with the External Solenoids menu. Each solenoid can be
Settings
connected to the following: Gas A, B, or C (or D, E, or F if the six gas
option is installed), the ozonator, or the permeation oven.
In the Main Menu, choose Instrument Controls > I/O Configuration >
External Solenoids.
MODE
DIAGS ALARM
MODE
DIAGS ALARM
Instrument Controls Menu
OPEN
<--
Model 146i Instruction Manual
Operation
3-59