Thermo Scientific 146i Instruction Manual page 91

Dynamic gas calibrator

Advertisement

Instrument ID
Communication Protocol
Thermo Fisher Scientific
BAUD RATE:
CURRENTLY:
SET TO:
OPER
MODE
The Instrument ID screen allows the operator to edit the instrument ID.
The ID is used to identify the instrument when using the C-Link or
MODBUS protocols to control the instrument or collect data. It may be
necessary to edit the ID number if two or more of the same instruments are
connected to one computer. Valid Instrument ID numbers are from 0 to
127. The Model 146i has a default Instrument ID of 51. For more
information about the Instrument ID, see Appendix B
Commands" or Appendix C
In the Main Menu, choose Instrument Controls > Communication
Settings > Instrument ID.
INSTRUMENT ID:
CURRENTLY:
SET TO:
OPER
MODE
The Communication Protocol screen is used to change the instrument
communication protocol for serial communications.
In the Main Menu, choose Instrument Controls > Communication
Settings > Communication Protocol.
9600
19200
?
CHANGE VALUE
SAVE VALUE
DIAGS ALARM
"MODBUS
Protocol".
51
52 ?
CHANGE VALUE
SAVE VALUE
DIAGS ALARM
Model 146i Instruction Manual
Operation
Instrument Controls Menu
"C-Link Protocol
3-49