C-Link Protocol Commands
Calibration
B-24
Model 146i Instruction Manual
Send:
set lamp 10.0
Receive:
set lamp 10.0 ok
lamp setting
This command reports the current photometer lamp setting. The example
below shows the lamp setting is 10.0%. This command is only available if
the photometer option is installed.
Send:
lamp
Receive:
lamp 10.0 %
o3 coef
This command reports O
instrument responds with "can't, wrong settings". The example below
reports that the O
coefficient is 1.000. This command is only available if
3
the photometer option is installed.
Send:
o3 coef
Receive:
o3 coef 1.000
set o3 coef value
This command sets the O
where value is a floating-point representation of the coefficient. The
example below sets the O
available if the photometer option is installed.
Send:
set o3 coef 1.005
Receive:
set o3 coef 1.005 ok
o3 bkg
This command reports the current O
reports that the O
background is 1.005 ppm. This command is only
3
available if the photometer option is installed.
Send:
o3 bkg
Receive:
o3 bkg 1.005 ppb
set o3 bkg value
This command is used to set O
value, where value is a floating-point representation of the background in
ppm. The example below sets the O
command is only available if the photometer option is installed.
Send:
set o3 bkg 15
Receive:
set o3 bkg 15 ok
coefficients. If the mode is incorrect, the
3
coefficients to user-defined values to value,
3
coefficient to 1.005. This command is only
3
background. The example below
3
background to user-defined values to
3
background to 15 ppb. This
3
Thermo Fisher Scientific