Table B-7. Solenoid Number Values - Thermo Scientific 146i Instruction Manual

Dynamic gas calibrator

Advertisement

C-Link Protocol Commands
Measurement Configuration
B-30
Model 146i Instruction Manual
set gas gas solenoid number
gas = | a | b | c | d | e | f |
number = 1-6
This command selects the solenoid number to be used when gas gas is
selected. The tank for the specified gas should be connected to the specified
bulkhead fitting. The example below shows that solenoid C is set for gas A,
according to
Table
B-7.
Send:
set gas a solenoid 1
Receive:
set gas a solenoid 1 ok
Table B–7. Solenoid Number Values
Value
Solenoid Value
1
Solenoid A
2
Solenoid B
3
Solenoid C
4
Solenoid D
5
Solenoid E
6
Solenoid F
sel gas gas span
gas = | a | b | c | d | e | f |
This command sets the gas gas to span levels 0, 1, 2, 3, 4, or 5. The
example below sets gas A to span level 1.
Send:
sel gas a span 1
Receive:
set gas a span 1 ok
gas gas span conc spanindex
gas = | a | b | c | d | e | f |
spanindex = span index 0-6 (1-5 reports the span concentration for each
span level and 6 reports the span concentration for manual mode)
This command reports the target span concentration for the specified gas
gas and the span level spanindex. The concentration reported is in ppb.
Send:
gas a span conc 1
Receive:
gas a span conc 1 123.45