Summary of Contents for Thermo Scientific Matrix 1029

  • Page 1 Thermo Scientific Matrix Electronic Pipette User Manual...
  • Page 2 Thermo Scientific Matrix Electronic Pipette User Manual Manual N09238...
  • Page 3: Table Of Contents

    CONTENTS Congratulations! ..................... 1 Before You Begin ..................... 1 A. INTRODUCTION ..................2 A.1 Thermo Scientific Matrix Single Channel Pipette ..........3 A.2 Thermo Scientific Matrix Multichannel Pipette ............3 B. PIPETTE OVERVIEW ................. 4 B.1 Components ......................4 B.2 General Description ....................5 B.3 Keypad ........................6 B.3.1 Operation Keys ....................6 B.3.2 Programming Keys ...................7 B.3.3 Special Function Key ..................8 B.4 Display ......................10 B.5 Pipette Specifications ..................11 B.6 Hazards and Precautions ..................12 C.
  • Page 4 D. CALIBRATION ..................23 D.1 Volume Measurement ..................23 D.2 Pipette Calibration .....................24 D.2.1 Factory Calibration Step ................24 D.2.2 “EASY” Cal Mode ...................24 D.2.3 “CALC” Mode ....................25 D.2.4 CALVIEW .......................27 D.2.5 CALRESET ....................27 D.3 Recommendations for Accurate Pipetting ............27 E. TROUBLESHOOTING ................29 F. MAINTENANCE ..................31 F.1 Single Channel Maintenance ................32 F.1.1 Single Channel Disassembly ................32 F.1.2 Single Channel Cleaning and Autoclaving ............34 F.1.3 Pipette Reassembly ..................34 F.2 Multichannel Maintenance .................35 F.2.1 Multichannel Disassembly ................35 F.2.2 Multichannel Cleaning and Autoclaving ............38 F.2.3 Multichannel Maintenance and Reassembly ..........38...
  • Page 6: Congratulations

    Congratulations! You have purchased the state-of-the-art in precision pipetting. The Thermo Scientific Matrix Electronic Pipette is the first cordless electronic pipette that “understands” the myriad of liquid handling protocols typically performed in today’s laboratory. Ergonomically designed and balanced to reduce fatigue, the Matrix electronic pipette functions so intuitively, you may need only a quick survey of this manual to use it. The electronic pipette executes your liquid pipetting tasks quickly, accurately, and with less risk of developing repetitive strain injuries (RSI) typically associated with manual pipette use. Laboratory results will become more standardized, while less time is expended on tedious bench procedures. Before You Begin Before using the Matrix electronic pipette, verify the contents of the package. The package should contain the following items: 1 Electronic Pipette 1 Operator’s Manual 2 Rechargeable NiMH Batteries or 1 Lithium-Ion Battery (16 Channels) 1 AC Power Transformer 1 Warranty Card 1 Accuracy/Precision Data Sheet If items are missing, contact a Thermo Fisher Scientific representative or contact customer service at 800-345-0206. Note: Batteries must be installed and recharged before using the Matrix electronic pipette Refer to the following sections in this manual: Section F.3, Installing/Replacing Batteries Section F.4, Charging/Recharging Batteries...
  • Page 7: Introduction

    A. INTRODUCTION The Thermo Scientific Matrix Electronic Pipette provides the convenience of a manual pipettor, with the programming capabilities of an electronic pipette. The Matrix electronic pipette features include: • Expanded memory for storing up to five pipetting programs. • Each program can hold a maximum of 40 steps. • “Scratch Pad” programming mode which allows for fast and simple program changes that are not stored into memory. • Enhanced speed control offers slower aspirating and dispensing of viscous liquids. Ideal for gel loading! • “EASY” CAL which simplifies in-lab calibration verification using distilled water. The Matrix electronic pipette is lightweight and cordless, utilizing rechargeable batteries, that can be used during recharge (with the exception of single channel pipettes). The pipette also has an automatic shut-off feature to extend battery life. The electronic pipette uses integrated circuit technology to store programming information. It will hold up to 40 steps in a program, with up to 5 programs saved in memory, and 1 scratch pad (See section C.1.1 Open Programming Mode or “Scratch Pad”). The program information is retained in memory even if the batteries are fully discharged or removed. The electronic pipette uses a “paced dispense” feature that performs repeat pipetting steps while the trigger is depressed.
  • Page 8: Thermo Scientific Matrix Single Channel Pipette

    Introduction A.1 Thermo Scientific Matrix Single Channel Pipettes The single channel electronic pipette are available in a number of model configurations (see Figures A-1) from 12.5 µl to 1250 µl, ranging in volume from 0.5 µl to 1250 µl. Each pipette is color-coded for easy volume identification (see section B.5, Pipette Specifications). Two models, 250 µl and 1250 µl, are offered with a shorter ejector sleeve reducing the overall length compared to the standard pipette. Ideal for use in confined spaces such as biosafety cabinet 1250 µl 250 µl 1250 µl 125 µl 30 µl 12.5 µl Shorty Shorty 250 µl Figure A.1: Single Channel Models A.2 Thermo Scientific Matrix Multichannel Electronic Pipettes The multichannel electronic pipettes are available in a number of configurations from 8 to 16 channnels. An expandable EXP pipette is...
  • Page 9: Pipette Overview

    Pipette Overview B. PIPETTE OVERVIEW B.1 Components Figure B.1a: Single Channel Components Figure B.1b: Multichannel Components...
  • Page 10: General Description

    Pipette Overview B.2 General Description Keypad. The keypad contains the operation keys that are used in programming the pipetting sequence and volumes. (See Section B.3, Keypad). Display. The display is an LCD screen that shows the pipetting operation (fill, dispense, mix, or purge) and fluid volume for each step in the pipetting program. (See Section B.4, Display). Battery Case. The battery case contains the rechargeable batteries. Nickel Metal Hydride batteries fully charge in 14-16 hours, while Lithium-Ion batteries fully charge in 2 hours. The pipette can perform a full day’s workload with no need to recharge. Battery life is extended with automatic shut-off after 10 minutes of inactivity. (See Section F.3, Installing/Replacing Batteries). Pipetting Trigger. The pipetting trigger initiates a program. A single pipetting step is executed by a quick press and release of the trigger. Repeat pipetting is executed by pressing and holding the trigger. Note: At the end of any dispense, the pipette tips should be removed from the fluid before releasing the trigger, to prevent aspiration of fluid.
  • Page 11: Keypad

    Pipette Overview B.3 Keypad The keypad contains the operation keys, programming keys, and the special function key. Figure B.3: Electronic Pipette Keypad B.3.1 Operation Keys [On/Off] The On/Off key turns the pipette on or off. When the pipette is turned on, it emits a beep tone, flashes “ZERO” on the display, then automatically resets the pistons to zero position. After zeroing, the display shows the first step of the program in the RUN mode. Note: If the pipette is being programmed for the first time, the first step will be the fill operation, with the default fill volume. (See Section C.1, Programming the pipette, for the default settings).
  • Page 12: Programming Keys

    Pipette Overview [Purge] The Purge key enters a purge prompt anytime during a pipetting operation to abort the program. By pressing the Purge key, then pressing the trigger, you can dispense any remaining liquid from the pipette tips and return to the beginning of the program. The Purge key is also used to enter the purge step into a program. Note: If there is insufficient liquid for a final dispense step, the pipette emits a beep tone and automatically displays the purge prompt. Note: The RUN mode performs the operation that is displayed, such as fill, dispense, mix, or purge.
  • Page 13: Special Function Key

    Pipette Overview Note: If you press and hold in the trigger during the final dispense step, the pipette performs the automatic blowout and holds its position until you release the trigger. This delay allows you sufficient time to remove the tips from the liquid to prevent aspiration of liquid as the pistons return to zero position. The Volume key controls the quantity of liquid to be aspirated or dispensed. The volume can be entered when the “Vol” prompt appears. Pipetting volume can be incremented quickly by continuously depressing the Volume key. The longer you hold the key, the greater the incremental speed. The Volume key can also be used to change pipetting speeds within a stored program or during the execution of a program. (See Section C.1.5, Adjusting Pipetting Speed). [Enter] The Enter key stores the program in memory. After each step is programmed, the Enter key is pressed once to accept the step. After the final step is programmed, the Enter key is pressed twice to denote the end of the program. The pipette verifies the end of the program with a triple beep tone. B.3.3 Special Function Key The Special Function key is the small hole below the Volume key. It can be activated with a pipette tip. By repeatedly pressing the Special Function key, you can display four functions: Pipetting Speed (SPd), Beep Tone (TONE), Paced Dispense Speed (PACE), and Calibration (CAL). To exit the Special Function key at any time, press the Enter key.
  • Page 14 Pipette Overview Pipetting speed can be adjusted by pressing the Volume key in the special function mode. Five speed settings are available: Speed Setting Display Slow Medium Slow Medium Medium Fast Fast TONE The beep tone signals the completion of one or more steps, or signals a warning: • Single beep tone: end of pipetting step. • Double beep tone: operation or programming key is not active in current mode; insufficient dispense volume, purge required; or illegal programming step attempted (see Section E, Troubleshooting). • Triple beep tone: end of pipetting sequence; end of programming sequence. There are three options to choose from in setting the beep tone. ON1 sounds all tones at the appropriate times listed above. ON2 uses only double and triple beep tones. OFF turns off all tones except for the tone to signify an illegal programming step. Press the Special Function key twice to display “TONE” and its current setting. To change the setting, press the Volume key. If the beep tone is set for ON1 or ON2, the beep tone symbol “ ” appears at the bottom of the display.
  • Page 15: Display

    Pipette Overview PACE The pace controls the speed of repeat pipetting (paced dispensing), FAST, SLOW or OFF. Press the Special Function key three times to display “PACE” and its current setting. To change the setting, press the Volume key. Paced Dispense. When using the paced dispense feature, the pipette dispenses the programmed volume at the specified pace (fast or slow). Immediately after each dispense step, the pipette emits a single beep tone (if beep tone is on) to prepare you for the dures). next dispense step (see Section C.5, Operating Proce The calibration of the pipette is factory set for distilled water at room temperature. If you are pipetting fluids with significantly different specific gravities or temperatures, use this setting to recalibrate the pipette. (See Section D, Calibration). The current speed setting appears at the bottom of the display. These settings can be selected and changed at any time before a program step is executed. B.4 Display The pipette display is a dynamic LCD screen. It displays the current operation, volume prompt (if entering a program), pipetting volume, beep tone symbol (if the beep tone is on), pipetting speed, and program number. Figure B.4: Pipette Display...
  • Page 16: Pipette Specifications

    Pipette Overview B.5 Pipette Specifications Item Volume Color Accuracy* Precision** Volume Channel Range Ring (±) 1029 12.50 µl Single Ch 0.50-12.50 µl 1.0%/0.1 ul 0.4%/0.06 ul 1020 30.0 µl Single Ch 1.0-30.0 iµl Purple 1.0%/0.15 µl 0.3%/0.05 µl 1021 125.0 µl Single Ch 2.0-125.0 µl Yellow 0.6%/0.3 µl 0.2%/0.1 µl 1022 250 µl Single Ch 5-250 µl Blue...
  • Page 17: Hazards And Precautions

    Pipette Overview B.6 Hazards and Precautions There are no known hazards associated with the Thermo Scientific Matrix Electronic Pipette when it is operated in accordance with the instructions in this manual. However, you should be aware of situations that can result in serious injury. Note: Do not perform troubleshooting procedures on the internal components unless instructed byThermo Fisher Scientific Technical Service personnel. WARNING! Ensure that the power transformer is connected to a power receptacle that provides voltage and current specified by Thermo Fisher Scientific. Use of an incompatible power receptacle can cause shock and fire hazard. CAUTION! Use only the power transformer supplied by Thermo Fisher Scientific. Use of an incompatible power transformer can damage the pipette. CAUTION! Always turn off the power and unplug the power transformer before cleaning the exterior. Fluid seepage can damage internal components. CAUTION! Do not immerse the unit in liquid. Fluid seepage can damage internal components.
  • Page 18: Programming And Operation

    Programming and Operation C. PROGRAMMING AND OPERATION C.1 Programming the Electronic Pipette gramming the Matrix electronic pipette is fast and easy. Pipetting steps and volume are entered in the sequence that they will be performed, then the program is saved to memory. This program will stay in memory (even if the pipette is turned off, or the electronic pipette battery runs low) until you change it. Five programs and one scratch pad (Program 0), each capable of holding 40 steps, can be entered and stored in the electronic pipette. C.1.1 Open Programming Mode or “Scratch Pad” (Program 0) The Matrix electronic pipette provides ample program storage for five, 40-step programs. In addition, it offers open programming mode, also known as the “scratch pad.” This mode is identified by “P-0,” which appears in the program section of the display (refer to the pipette display on page 10). This feature enables quick changes to pipetting protocols without affecting previously stored programs. To use the “scratch pad” mode: From the start of any program, simply begin entering a new protocol (selecting a pipetting operation as described below). This new protocol can now be used, and will remain active until a stored program is retrieved. “P-0” appears in...
  • Page 19 Programming and Operation Fill Dispense Mix Volume Volume Volume Volume 12.50 10.00 10.00 5.00 30.0 25.0 25.0 10.00 125.0 100.0 100.0 50.0 250.0 250.0 50.0 50.0 1250 1250 Table C.1.2: Default Programming Volumes (in µl) 2. Select pipetting operation. Enter the programming mode by pressing the [Fill] or [Mix] key.
  • Page 20 Programming and Operation Mix. The [Mix] key programs the pipette to perform a combination of fills and dispenses. You can program a mix volume or mix cycle (single fill and dispense). You can press [Mix] once to program a mix volume, or twice to program a mix cycle. Mix cycles from 2 to 9 are available on the pipette. The default setting is three mix cycles. Some programming considerations for the mix operation: • If the program begins a mix cycle from zero position, it will complete the cycle with an automatic blowout, then display “ZERO.” Press the trigger to return the pistons to zero position. • If the program begins a mix cycle above zero position (volume is greater than zero), the fluid is dispensed, and the mix cycle begins. The mix cycle will end at the same volume it started with. Also, the mix operation will remain in the display so it can be repeated (useful for serial dilutions), unless followed by another program step. To end the mix operation, press the [Purge] key, then press the trigger. Purge. The [Purge] key programs the pipette to purge the fluid. This step can be entered as a program step, or it can be pressed any time while pipetting to terminate the program and dispense the remaining liquid. Note: If the program ends at a volume greater than zero, and the last step is not a mix cycle, the “Purge” prompt automatically appears. Also, the pipette emits a double beep tone to warn you that some fluid remains.
  • Page 21: Storing A Program

    Programming and Operation Confirm programming step. Press [Enter] to confirm the programming step. The display flashes all of the operation keys to prompt the next programming step: Figure C.1.2b: Flashing Display 5. Select next pipetting operation. Repeat steps 2 to 4 until all programming steps have been entered. 6. Save program and exit. When the display flashes all operation keys, press [Enter] to save the program and exit the programming mode. The pipette emits a triple beep tone to signal the end of the programming sequence. The first step in the program is displayed. “Run” is displayed in the lower left corner to indicate that the new program is ready to be run. Go to Section C.1.6 for Reviewing the Program, or Section C.3 for Operating Procedures. C.1.3 Storing a Program To create and store programs in the pipette: 1. Select the program number by pressing the [Prog] key, then press the [Vol] key until the desired program number (P1-P5) is displayed. 2. Press and hold the [Prog] key, while simultaneously pressing the Special Function key to enter the programming mode. The prompt “Prog” will appear on the display. 3. Begin entering your new program (refer to Section C.1.2, Entering a Program in Open Programming Mode, for instructions).
  • Page 22: Program Recall

    Programming and Operation The program number will be displayed for two seconds. (The first step of the newly written program is displayed). This program is ready to be run. C.1.4 Program Recall To recall a stored program: 1. Press the [Prog] key, then the [Vol] key to select the desired program number. 2. Press the [Enter] key to activate the displayed program. C.1.5 Adjusting Pipetting Speed The Matrix electronic pipette offers unique speed control programming for Fill, Dispense, Mix, or Purge. Speeds within any of the five stored programs can be independently set and stored as an integral part of the pipetting protocol. Pipetting speed is easily changed at the start of each pipetting function. To change pipetting speed during the execution of a program: Press the [Vol] key to select the speed for each pipetting step (Fill, Dispense, or Mix only). If purge is programmed as a step, its speed can also be programmed. Note: Purge must be changed in the program. This newly set speed will remain active with each function even if the pipette is turned off. Speed settings within any of the five stored programs will also remain in memory with the program functions until changed by the operator. C.1.6 Reviewing the Program After programming the pipette, review the programming steps: 1.
  • Page 23: Programming Examples

    Programming and Operation 3. Observe display. At each step of the program, observe the display. If necessary, re-enter the programming sequence. Some helpful hints for reviewing your program: • If the program is complex and contains several steps, write the steps on paper before entering the program. Check off each step as you review the program. • If the program is intended for paced dispensing, review the program again, with the trigger depressed until the last step. Check the dispense pace to ensure the dispense steps are a comfortable speed. C.2 Programming Examples Examples of different pipetting sequences are described below, and are ideal for use as training exercises. Press the keys shown in brackets [ ], then press the [Vol] key until the volume in parentheses ( ) is displayed. At the end of the program, a triple beep tone is sounded. (see section D.3 Recommendations for Accurate Pipetting for suggestions for better precision and accuracy). C.2.1 Sample Transfers Example: Fill and dispense 100 µl. Option 1 Option 2 [Fill]...
  • Page 24: Incremental Pipetting

    Programming and Operation C.2.2 Incremental Pipetting Example: Fill and dispense 250 µl. Note: The 10 µl Primer Dispense reduces the percentage of error in [Fill] (250) [Enter] incremental dispensing (see section D.3 Recommendations for Accurate Pipetting) [Disp] (10) [Enter] [Disp] (45) [Enter] [Enter] In this program, only one dispense step needs to be programmed. If any program ends in a dispense step, the pipette repeats the dispense step until the volume is zero, or is less than the programmed dispense volume. If the remaining volume is less than the programmed dispense volume, the pipette displays a “Purge” prompt to discard the remainder. This discarded volume contains the cumulative error of previous pipettings (see section D.3 Recommendations for Accurate Pipetting for further explanation).
  • Page 25: Simple Dilutions

    Programming and Operation C.2.4 Simple Dilutions Example: Prepare 1:10 dilution with a 10 µl sample volume. [Fill] (100) [Enter] [Fill] (20) [Enter] [Fill] (10) [Enter] [Purge] [Enter] [Enter] The first fill volume is for the diluent (100 µl), followed by an air gap (20 µl), and finally by the sample volume (10 µl). The purge step will dispense the total volume. C.2.5 Supernatant Collection Example: Aspirate 100 µl from 8 rows of a Microtiter plate and discard the ® total volume. [Fill] (100) [Enter] [Fill] [Enter] [Fill] [Enter] [Fill] [Enter] [Fill] [Enter] [Fill] [Enter] [Fill] [Enter] [Fill] [Enter] [Purge] [Enter] [Enter] This program is performed with the 850 µl handle. All 100 µl fill volumes are collected in the same tips. When completed, the total volume is discarded with the purge step.
  • Page 26: Operating Procedures

    Programming and Operation C.3 Operating Procedures After programming the pipette, the display will show the first program step, with the “Run” prompt displayed in the lower left corner: Figure C.3: Run Prompt The following steps describe a basic program for pipetting reagent into a microtiter plate. The pipette is programmed as follows: Note: The 10 µl [Fill] (250) [Enter] Primer Dispense reduces the percentage of error in incremental [Disp] (10) [Enter] dispensing (see section D.3 Recommendations for [Disp] (30) [Enter] [Enter] Accurate Pipetting) 1.
  • Page 27 Programming and Operation 4. Dispense liquid. Position the tips over the first column of wells in the microtiter plate, and press the trigger to dispense 30 µl. At the completion of the dispense step, the pipette emits a single beep tone (if the beep tone is set to ON1). 5. Repeat dispense. Repeat the previous step for the rest of the columns in the microtiter plate. At the completion of the eighth dispense, the pipette emits a double beep tone (if the beep tone is on, otherwise it emits a single beep tone) and displays the “Purge” prompt. There is insufficient volume for another 30 µl dispense, and the pipette signals for a purge step. -or- Paced dispense. Steps 4 and 5 can also be performed as a paced dispense. Instead of pressing the trigger for each dispense step, press and hold the trigger to automatically dispense each volume. Immediately after each dispense step, the pipette emits a single beep tone (if the beep tone is on), prompting you to prepare for the next dispense step.
  • Page 28: Calibration

    Calibration D. CALIBRATION Each electronic pipette is factory-calibrated for distilled water at 20 °C. For easy calibration using distilled water, the pipette features an “EASY” CAL mode (see Section D.2.2, “EASY” CAL mode). When pipetting liquids of significantly different specific gravities or temperatures, however, the Calibration (CALC) feature may be used to achieve greater accuracy (see Section D.2.3, “CALC” mode). To prepare for calibration, ensure the pipette and tips are equilibrated. When calibrating a multichannel pipette, select only one channel to pipette with. Using an analytical balance, first obtain the actual value of the programmed volume at factory calibration. The programmed volume (Vp) used is typically the calibration point for that pipette (Refer to Table D.2.2) D.1 Volume Measurement To measure the actual value of the programmed volume: Fill the programmed volume (Vp), then dispense the entire volume into a container on the analytical balance. 2. Read the weight on the analytical balance. Repeat this procedure ten times. ote: Use a clean and dry pipette tip with each pipetting cycle. When filling the liquid, immerse the pipette tip approximately 1/4 inch (approx. 0.6 cm) below the liquid surface.
  • Page 29: Pipette Calibration

    Calibration D.2 Calibration Electronic pipette users can calibrate their pipettes with “EASY” CAL mode for distilled water or with “CALC” mode to calibrate other liquids and for temperatures other than room temperature. D.2.1 Factory Calibration Step The factory calibration step is needed when calibrating your electronic pipette. This is located on the calibration certificate called “Pipette Calibration Report” enclosed in the documention for this pipette. (See comments on page 1 of Pipette Calibration Report). View the pipette’s current calibration step by using the CALVIEW mode listed in section D.2.4. Note: When resetting the calibration step to a default value, as listed in section D.2.5 CALRESET, the step is set to an average value for all pipettes of its class. This value may differ from the factory-optimized calibration step documented on the “Pipette Calibration Report”...
  • Page 30: Calc" Mode

    Calibration 1. Weigh out ten aliquots of distilled water on an analytical balance using one pipetting channel. The programmed dispense volume must correspond to the calibration point for that pipette (for example, 200 µl for a 250 µl pipette). Refer to Table D.2.2, Calibration Specifications, for proper calibration points. Note: Ensure that the distilled water used is at room temperature (20-22.5 °C). An internal correction factor of 1.003 is used by the pipette to convert weight (gm) to volume (ml).
  • Page 31 Calibration 2. Calculate display volume of the pipette: D v = x V p = Programmed Volume V a = Corrected Actual Volume (corrected for temperature) C p = Calibration Point (each pipette handle has a unique calibration point, see table D.2.2: Calibration Specifications.) D v = Display Volume 3. Enter the CAL mode on the pipette by pressing the Special Function key (use a pipette tip to press the key) until “EASY” appears on the display. 4. Press the [Vol] key to display “CALC” (pressing the [Vol] key again will return you to the “EASY”...
  • Page 32: Calview

    Calibration D.2.4 CALVIEW CALVIEW displays the pipetting stroke (number of motor steps) used to reach the calibration point. This value allows you to confirm the calibration during pipetting and after calibration. To activate CALVIEW, enter the CAL mode using the Special Function key. When the calibration point is displayed, exit press the [Fill] key. The number of steps is displayed. Press any key to CALVIEW. D.2.5 CALRESET CALRESET allows you to return to a default calibration step. To reset the calibration, enter the CAL mode using the Special Function key. When the calibration point is displayed, press the [Purge] key. The display flashes “FACT,” “CAL,” then “SET” and the pipette resets to the default step. This default setting is an average value for all pipettes of its class. This value may differ from the factory-optimized calibration step documented on the “Pipette Calibration Report” certificate enclosed in this pipette’s documentation. D.3 Recommendations for Accurate Pipetting Matrix Electronic pipettes are factory calibrated to deliver accurate volumes with distilled water at 21 °C and 50% humidity. By using the calibration procedures described above, you can pipette liquids of different specific...
  • Page 33 Calibration • Calibrate the pipette for the liquid. If the liquid has a specific gravity and vapor pressure equivalent to distilled water, calibrate using “EASY” CAL mode. Otherwise, use “CALC” mode. • Pre-wet the pipette tip. Aspirate and dispense a few times before aspirating the liquid to increase humidity within the tip, reducing sample evaporation that results in lower dispense volumes in initial few pipettings. The need to pre-wet increases when working with liquids with high vapor pressure and/or high surface tension. • Immerse tips fully. Tips should be immersed approximately 1/4 inch (approx. 0.6 cm) below the liquids meniscus and away from the container surfaces to avoid aspiration of air and excess droplets clinging to the outside of the tip. • Discard a primer volume. When the motor changes from aspiration to dispense, the percentage of error is greater in the first dispense than in subsequent dispenses. • Pause consistently. It takes a moment for the liquid in the tip to finish moving. Pause with the tip in the liquid for about one second after aspirating the sample to allow the liquid to finish moviing into the pipette tip after the pipette stops. Otherwise, the volume will be lower than it should.
  • Page 34: Troubleshooting

    Troubleshooting E. TROUBLESHOOTING Possible procedural and system problems, as well as recommended solutions. If you cannot resolve the problem, call technical service at 800-345-0206. No power to the pipette. Probable Cause Recommended Solution Recheck the orientation of the batteries; Batteries are incorrectly they should match the diagram in the installed. battery case Ensure the connections to the battery pack Battery pack cables are not are securely plugged into the connection properly connected. from the pipette body. Battery Spacer is not in Ensure that the battery spacer is in place unit or is not in correctly for all models that use Nickel Metal (Multichannels only with NiMH). Hydride (see section F.3.1). Double beep tone occurs.
  • Page 35 Troubleshooting The motor sounds rough and aspiration is very slow. Probable Cause Recommended Solution Verify speed setting, the slowest speed The pipette has been set to the setting normally causes the motor to run slowest speed. slowly and louder than usual. Motor has not been used for an Cycle the unit (fill and dispense) five to extended period. seven times to re-lube the piston. After reassembly, the pipettor leaks or does not fill liquid. Probable Cause Recommended Solution Disassemble the pipette and reseat the O-ring was not properly seated.
  • Page 36: Maintenance

    Maintenance F. MAINTENANCE The Thermo Fisher Matrix family of electronic pipettes has been specifically designed for serviceability, enabling the user to service instruments with relative ease. Under certain circumstances where on-site service is not feasible, the instrument may need to be returned to Thermo Fisher Scientific for repair (see Pipette Service, Sections G.2 and G3). Electronic pipettes require very little maintenance. The exterior of the pipette can be cleaned periodically with a soft cloth moistened with methyl alcohol. Periodic cleaning of the tip fitting may be necessary using the procedures for maintenance listed below. In cases where pipettes have been exposed to biohazardous materials, autoclaving internal components is required. Thermo Fisher Scientific provides the following Preventive Maintenance/Autoclave Kit: Preventive Maintenance/Autoclave Kit • Replacement cross tubes • O-ring lubricant • Wooden applicator • O-rings Item # Description Item # Description 9809 12.5µl - 6, 8, & 12 channel 9804 850µl - 12 channel 9801 125µl - 6, 8, & 12 channel 9805 1250µl - 6 & 8 channel 9802 250µl - 6, 8, & 12 channel CAUTION! DO NOT AUTOCLAVE entire unit or clean the keypad with bleach solution or other solvents. Extreme heat can damage the display and other electronic components. Caustic cleaning solutions can damage the keypad.
  • Page 37: Single Channel Maintenance

    Maintenance F.1 Single Channel Electronic Pipette Maintenance F.1.1 Single Channel Electronic Pipette Disassembly With proper use of the pipette, the disassembly and cleaning of the pipette tip is not necessary. However, if fluid is inadvertently drawn into the tip fitting, then the lower end of the pipette should be disassembled for thorough cleaning. To disassemble the lower end of the pipette: 1. Detach the collet by twisting it counter-clockwise (see Figure F.1.1a). 2. Remove the collet and ejector sleeve by sliding both pieces down over the cylinder (see Figure F.1.1b). For reassembly later, note the position of the large spring in the collet. Figure F.1.1a: Detaching the Collet 3. Unscrew cylinder counter-clockwise and remove it from the pipette body.
  • Page 38 Maintenance Figure F.1.1c: Pipette Tip Assembly...
  • Page 39: Single Channel Cleaning And Autoclaving

    Maintenance F.1.2 Single Channel Cleaning and Autoclaving General Cleaning: Carefully wipe down all disassembled plastic and metal parts with methyl alcohol or 70% isopropyl alcohol. Decontamination/Autoclaving: Decontaminate all disassembled parts by wiping them with a bleach solution (5% sodium hypochlorite) or by autoclaving the parts at 121 °C for 20 minutes. Re-lubrication: Lightly coat piston with O-ring lubricant and replace or re-install O-ring on the piston, as described in reassembly procedure. Note: Do not clean the O-ring. Replace the O-ring after decontamination or autoclaving (O-ring and O-ring lubrication is available from Thermo Fisher Scientific). F.1.3 Pipette Reassembly After cleaning or decontaminating the disassembled parts, the pipette can be reassembled (refer to Figure F.1.1c for parts identification and reassembly...
  • Page 40: Multichannel Maintenance

    Maintenance 4. Slide the collet, with spring and ejector sleeve, over the cylinder. Position the metal tab of the ejector trigger behind the tab on the ejector sleeve (see Figure F.1.3b). 5. Attach the collet onto the pipette body by twisting left slightly, then right, to insert the tabs of the pipette body into the locks on the collet (see Figure F.1.3b). 6. Test the ejector function to ensure that the ejector sleeve moves down when the ejector trigger is pressed. Figure F.1.3b: Detaching the Collet F.2 Multichannel Electronic Pipette Maintenance F.2.1 Multichannel Electronic Pipette Disassembly 1. Remove the color ring by gently prying loose one of the edges near the handle trigger.
  • Page 41 Maintenance 4. Remove plastic ejector trigger with spring and slide the aluminum ejector comb off of the tip fittings (see Figure F.2.1b). On 16 channel models, reverse this process and remove comb first, then trigger. 5. Remove the piston/ cylinder assembly by pulling the motor screw down from the motor shaft coupling and out of the main housing. Remove the manifold assembly by firmly lifting it out of the main housing support ribs Figure F.2.1b: Removing Ejector (see Figure F.2.1c). Trigger and Ejector Comb Figure F.2.1c: Removing Piston/Cylinder and Manifold Assemblies...
  • Page 42 Maintenance Item Description Item Description Keypad Tip fitting Main board Console cover Flex circuit Console display Battery box Batteries Driver board Main housing Motor Trigger hinge pin Color ring Handle trigger Motor coupling Handle trigger spring Motor screw Ejector trigger Piston backplate Piston Photosensor O-ring Cylinder Cross tube Cover housing Manifold plate Housing screw Figure F.2.1d: Pipette Components...
  • Page 43: Multichannel Cleaning And Autoclaving

    Maintenance F.2.2 Multichannel Cleaning and Autoclaving 1. Remove the piston assembly from the cylinder. 2. Remove and discard the cross tubes into a biohazardous container. 3. Remove the O-rings (only if necessary, see section F.1 and/or F.2 for O-ring Replacement) and discard them into a biohazardous container. 4. Place all components (see Figure F.2.2) into the autoclave and autoclave at Figure F.2.2: Components to be Autoclaved 121°C for 20 minutes. F.2.3 Multichannel Maintenance and Reassembly Once components are autoclaved and completely dry, use the following procedures to relubricate and reassemble the unit: O-ring Replacement The O-rings may require replacement if liquids have been over aspirated. In this case, the O-rings should be removed prior to autoclaving. Spare O-ring...
  • Page 44 Maintenance 2. Install new O-rings. After the old O-rings have been removed and all components have been autoclaved, the new O-rings may be installed. Before installing the O-rings, apply a thin coat of lubricant to each O-ring, using the O-ring lubricant supplied in the Preventive Maintenance/ Autoclave Kit. Figure F.2.3a: Removing O-rings Re-lubrication • 1250 µl, 850 µl, 250 µl Pipettes: Use the wooden applicator and O-ring lubricant supplied in the Preventive Maintenance/Autoclave Kit to apply a thin layer of lubrication approximately 0.5 inches (approx. 1.3 cm) down into each piston hole that requires it. Spread the lubricant evenly along the inner wall of each cylinder by using a circular motion with the wooden applicator (see Figure F.2.3b). Figure F.2.3b: Re-lubricating 1250 - 250 µl Pipettes • 125 µl, 30 µl, 12.5 µl Pipettes:...
  • Page 45 Maintenance Reassembly 1. Reinsert the piston assembly into the lubricated cylinder. Ensure that the motor screw engages into the hole in the center of the cylinder (see Figure F.2.3d). NOTE: To correctly orient the piston and cylinder assembly, ensure that the lot number of the cylinder is facing up and the painted section of the piston backplate is facing down. Also, care should be taken to keep the O-rings from being pinched when inserting the pistons into the cylinder.
  • Page 46 Maintenance Tip Fitting Positions Cylinder Positions 16 Channel 12 Channel 8 Channel 6 Channel 5 Channel 8 Channel 6 Channel 5 Channel Table F.2.3b: Cylinder/Cross Tube Positions Note: If the Cross Tube orientation between the cylinder and Tip Fittings is incorrect, performance of the pipette will be impaired. 3. Carefully insert the flat end of the motor screw into the hole of the motor coupling. This can be accomplished by holding the piston/cylinder assembly at a slight angle while inserting the motor screw into the hole...
  • Page 47 Maintenance Figure F.2.3e: Installing Piston/Cylinder Assembly 4. After installing the piston/cylinder assembly, install the manifold assembly into the main housing. To keep the tubing free from ejector contact, rotate the manifold plate and tubing one complete turn counterclockwise (see Figure F.2.3f). Insert the manifold assembly into the main housing support ribs. Figure F.2.3f: Rotating Manifold Assembly 5. Slide the ejector comb over the tip fittings with the slots on the back of the comb exposed (see Figure F.2.3g). Figure F.2.3g: Attaching Ejector Comb...
  • Page 48 Maintenance 6. Hold the ejector comb in place while inserting the ejector trigger (see Figure F.2.3h). Place the end of the ejector spring against the plastic stop and compress the spring while positioning the ejector trigger over the aluminum ejector comb (the ribs on the ejector trigger should mate with the slots on the aluminum ejector comb). For 16-channel models, install the trigger prior to connecting the ejector comb. Figure F.2.3h: Inserting Ejector Trigger NOTE: Be careful not to pinch the cross tubes when inserting the ejector trigger.
  • Page 49: Installing/Replacing Batteries

    Maintenance F.3 Installing/Replacing Batteries The Matrix electronic pipette comes with rechargeable Nickel Metal Hydride (NiMH) batteries or with a Lithium Ion battery for 16 channel models. Under normal use, the batteries can be recharged approximately 400-500 times. When they can no longer hold their charge, the batteries must be replaced. Replacement batteries can be purchased through customer service. The pipette may also be returned to Thermo Fisher Scientific for battery replacement. Replacement batteries: 6, 8, 12 Multichannels 2 NiMH Item No. 6092 Single Channel 1 NiMH Pack Item No. 2092 16 Channel only 1 Lithium-Ion Battery Item No. 2094 Note: When replacing batteries, ensure the pipette is turned off before removing the battery case. F.3.1 Installing/Replacing Nickel Metal Hydride Batteries F.3.1A Installing/Replacing Single Channel Electronic Pipette Batteries 1.
  • Page 50 Maintenance 3. Disconnect wire connectors. Pull apart the two wire connectors over the battery pack (see Figure E.3.1Ab). 4. Remove battery pack. Pull gently on the battery wire connector (without the pins) to dislodge the battery pack, then lift the battery pack out of the main housing (see Figure F.3.1Ac). Figure F.3.1Ab: Disconnect Wires 5. Insert new battery pack. Insert the new battery pack into the cavity of the main housing. Ensure that the wire is looped under the battery pack and the connector is above it (refer to Figure F.3.1Ac). NOTE: The battery pack fits very snugly in the main housing;...
  • Page 51 Maintenance 7. Reassemble pipette top. Attach pipette top to the main housing. Ensure flex cable is behind the top of the pipette trigger and the trigger contacts the limit switch properly (see Figure F.3.1Ae). Fit the edges of the pipette top snugly within the edges along the top of the pipette housing. Ensure that the cables to the power jack are inside the main housing well below the edges to avoid pinching the cables. Hold assembly together firmly while resecuring the two Figure F.3.1Ae: screws. Assembling Pipette Top 8. Charge battery. Charge the new battery before using the pipette (see Section F.4, Recharging Batteries). F.3.1B Installing/Replacing Multichannel Electronic Pipette Batteries The Matrix Electronic pipette is shipped with Nickel Metal Hydride batteries packaged separately. Install the batteries and recharge them for 14-16 hours.
  • Page 52 Maintenance 1. Remove battery case. Position your thumb over the thumb grips (see Figure F.3.1Ba) and brace your fingers against the front of the keypad. Apply firm pressure to the panel with the thumb grips to release the latch. 2. Remove batteries. Press a pipette tip against either battery through the slot in the bottom of the battery case to remove the batteries (see Figure F.3.1Bb). 3. Install new batteries. Insert the spacer into the bottom of the battery box (as shown in Figure F.3.1Bc). Insert Figure F.3.1Bb: Removing the Batteries new batteries by placing the positive end into the battery case first, then the negative end in place over the spacer (an exploded view of the assembly is shown in Figure F.3.1Bc, or the diagram in the battery case). Ensure that the batteries are correctly oriented.
  • Page 53: Installing/Replacing Lithium-Ion Battery (16 Ch)

    Maintenance F.3.2 Installing/Replacing Lithium-Ion Battery (16 Channel pipette) Your electronic 16 channel pipette is shipped with a Lithium-on battery. Before using the pipette, charge it for 2 hours. Use the following procedure when installing a new battery pack. Otherwise, continue to Section F.4, Recharging Batteries. 1. Remove battery case. Position your thumb over the thumb grips (see Figures F.3.2a,b) and brace your fingers against the front of the keypad. Apply pressure to the panel with the thumb grips to release latch. Carefully slide battery case out of the pipette, allowing sufficient slack for attached wires of the battery. Figure F.3.2a: Figure F.3.2b: Battery Case (Closed) Battery Case (Open) 2. Remove Console Cover. Unclip the two snaps on console cover assembly, then unhook the end of console cover assembly (see Figures F.3.2c,d). Figure F.3.2c: Unclip Snaps Figure F.3.2d: Unhook Cover...
  • Page 54 Maintenance 3. Carefully Remove Console Cover. Press and lift off console cover assembly (see Figure F.3.2e). BE CAREFUL not to tear the attached flex cable! (Refer to Figure F.3.2f). 4. Unclasp Battery Connector. Lift one side Figure F.3.2e: Remove Console Cover of battery connector with a small flathead screwdriver. Grasp all three wires and pull out connector (see Figures F.3.2f,g). CAUTION: Be careful not to tear the Flex Cable Figure F.3.2f: Lift Battery Connector Figure F.3.2g: Unclasp Battery Connector...
  • Page 55: Charging/Recharging Batteries

    Maintenance 5. Connect New Battery. Attach the wires of the new battery in place of the old battery. 6. Reassemble Console Cover. Attach the pipette top to the main housing. Fit the edges of the pipette top snugly within the edges along the top of the pipette housing. F.4 Charging/Recharging Batteries F.4.1 Charging/Recharging Nickel Metal Hydride (NiMH) Batteries The electronic pipette uses two Nickel Metal Hydride (NiMH) rechargeable batteries, located in a battery case under the keypad. The batteries are charged in the pipette by using the power transformer included in the package, with the exception of single channel pipettes which do not have this ability. You can operate the pipette while it is recharging, or you can operate it solely with the power transformer (without batteries). The batteries can be recharged approximately 400-500 times. To extend battery life, the pipette will turn off automatically if it is not used within ten minutes. 1. Attach transformer. Attach the cable end of the transformer into the pipette. Plug the transformer into the electrical outlet. 2. Recharge batteries. Recharge batteries for 14-16 hours to ensure a full charge. After the batteries have been recharged, unplug the pipette from the transformer. Note: Batteries hold their maximum charge if they are fully discharged before recharging again.
  • Page 56: Charging/Recharging Lithium-Ion Battery (16 Ch)

    Maintenance CAUTION! Avoid excessive charging of the batteries when the pipette is not in use. Excessive charge to the batteries will shorten the battery life and may damage the batteries. F.4.2 Charging/Recharging Lithium-Ion Battery (16 Channel Electronic Pipette) The 16 channel electronic pipette uses a Lithium-Ion rechargeable battery located in a battery case under the keypad. The battery is charged while it is in the pipette by using the power transformer included in the package. The pipettor can be operated while it is recharging. The battery can be recharged approximately 400-500 times. To extend battery life, the pipette will turn off automatically if it is not used within ten minutes. Rapid Charge - The battery should be charged for 1.5 to 2 hours to achieve a full charge. Refer to the following table for partial charge times: Amount of Charge Time 1.25 hour 1.5 hours 100% 2 hours max The battery pack for the 16 channel Electronic pipette features an indicator light to indicate the state of charge of the battery while plugged into the power supply. With the unit plugged in, the indicator operates as follows: • Green - Indicates that the power supply is connected. •...
  • Page 57 Maintenance As the unit continues to charge, current is supplied in pulses. This is indicated by the flashing of the indicator. As the battery nears full charge, the flashing slows down becoming nearly a constant green. When the battery is fully charged, the indicator will be almost a constant green, flashing only about once every 3-5 seconds to ensure the battery stays fully charged. Note: The typical life expectancy of the battery is approximately 400-500 full recharges. To prolong this and attain the maximum usage from the battery, it is recommended the battery be recharged whenever it is not being used instead of waiting until the battery is completely discharged.
  • Page 58: Technical Service

    G. TECHNICAL SERVICE G.1 Instrument Warranty This instrument is warranted to the original purchaser by Thermo Fisher Scientific to be free of defects in materials or workmanship for a period of 12 months from the date of purchase, providing that is has been operated according to the instructions, not abused or misused, that the serial number has not been removed, and that the instrument has not been disassembled, other than for procedures for normal maintenance. No other warranty is expressed or implied, including the warranty of merchantability. Should a unit need to be returned, the purchaser must pay insurance and shipping charges. Thermo Fisher Scientific will assume the cost of returning the repaired unit to the purchaser. This products packaging is capable of withstanding normal shipping hazards. Contact Thermo Scientific Technical Service for recommended equipment decontamination methods (see also section G.4), or to obtain repairs or replacements within the term of the above warranty at 800-345-0206 fax: (603) 577-7682. For Europe and Asia locations, please visit: www.thermo.com/matrix...
  • Page 59: Expert Pipette Service

    G.2 Expert Pipette Service Trusted Service Certified technicians calibrate and repair Thermo Scientific Matrix pipettes as well as most other pipette brands and models. With more than 20 years experience manufacturing and servicing pipettes, we offer a broad range of services to fit every need: basic calibrations, preventive maintenance and repair services.
  • Page 60: On Site Pipette Service Options

    Technical Service G.3 On-site Pipette Service Options Services include: • Physical pipette inspection • Leak test • Replacement of seal and O-ring, as needed • Gravimetric performance validation using NIST-traceable balance • Performance validation of five aliquots at two volumes •...
  • Page 61: Recommended Equipment Decontamination Methods

    Technical Service G.4 Recommended Equipment Decontamination Methods The following table lists the proper decontamination procedures for returning Matrix Electronic Pipettes to Thermo Fisher Scientific for service. Contact customer service at 800-345-0206 for a current equipment decontamination verification form. Heat at Autoclave 65 °C/ 70%/30% Ethylene 121 °C/ Product/Component 10 Hr. IPA/Water Bleach Oxide 15 min. Multichannel Pipettes Pipette Exterior YES Wipe Wipe YES Piston/Cylinder Assembly YES YES YES YES Tip Manifold YES YES...
  • Page 62 Technical Service...
  • Page 63: Tips And Reservoirs

    Technical Service G.5 Tips and Reservoirs The tips and other consumables you use play a role that’s equally as important as your instruments. Innovation at Thermo Fisher Scientific doesn’t stop with pipettes. Non Sterile Pipet Tips Pipettes Bulk Sterile Sterile Filtered ECOTIPs 30µl TallTip 12.5 µl / 30 µl 7600 7631 7632 7635* 250 µl Tip 125 µl / 250 µl 7250 7151 7152 7275** 7156 250 µl TallTip 125 µl / 250 µl 7280 7281 7282 7285** 300 µl Tip 125 µl / 250 µl 7320 7321 7322...
  • Page 64 Index INDEX Autoclaving 34, 38 Maintenance (Multichannel) 35-43 Automatic blowout Maintenance (Single channel) 32-35 Maintenance Procedures 31-52 Mix Batteries, Installation (Lithium-Ion) Mix Key Batteries, Charging (Lithium-Ion) Batteries, Charging (NiMH) Batteries, Installation (NiMH) On/Off Key Battery Case Open Programming Mode 13-16 Operation Keys CAL “CALC” Mode PACE Calibration (Procedures) 23-27 Paced Dispense 9, 22 Calibration (“CALC” Mode) Pipetting Trigger Prog Key Calibration (“EASY” Cal) Calibration (Specifications) Program (Examples) 18-20 CALRESET Program (Recall) CALVIEW Program (Storing) Color Ring Program Keys...
  • Page 66 Thermo Fisher Scientific, Inc. 81 Wyman Street P.O. Box 9046 Waltham, Massachusetts 02454-9046 United States Thermo Fisher Scientific Matrix Liquid Handling Products 22 Friars Drive Hudson, New Hampshire 03051 United States Technical Support: North America Tel: 800.345.0206 Email: matrix.technicalservice@thermofisher.com Hours of Operation: 8:00 AM to 5:00 PM ET Europe Tel: +44(0) 161 486 2112 Email: matrixtechcorp.eu.service@thermofisher.com...