Guide Holes and Capillary Clips
5.7.3

Installing the Partition Panel Plugs

Installing the plug with slit
Page 62
Guide holes and capillary clips are provided at specific positions on the
system modules. Route flow connections from one module to the next
module in the Vanquish system through the appropriate guide hole or
capillary clip when instructed to do so in the manual.
There are two types of partition panel plugs available in the detector
ship kit.
Figure 17: Plugs available for the partition panel
No.
Description
1
Plug with slit, for guiding capillaries with small outer diameter, such as
uninsulated capillaries.
2
Rotating plug, for guiding capillaries that do not fit in the plug with slit,
including insulated capillaries.
1. On the detector partition panel, push the plug with slit in the
required recess of the partition panel (if not present yet).
2. To secure the plug with slit, insert the nose into the opening of the
partition panel.
Figure 18: Securing the partition panel plug with slit