10 Installation: Ceph arm
10.3 Installation
96
9. Use the arrow keys (I) on the control panel to lower the unit until the
bearing bolts on the unit are at the same height as the drill holes on
the ceph support arm.
NOTICE! Check the room height before you raise the unit. If the room
height is less than 2.27 m (89 3/8") or 2.30 m (90 1/2") for installation
with the floor stand, you must limit the maximum travel height (S018:
Test step 2 [ → 149]).
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016