Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
S017: Test step 18
11.12.3.9
Occlusal bite block function
The occlusal bite block function can be activated or deactivated via
service routine S017.18.
Should the customer wish to set a nominal angle deviating from the
factory setting (7°), make the corresponding configuration settings via
service routine S018.7).
Selection field
1
* factory settings
1. Call service routine S017.18 [ → 147].
 Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use arrow keys (A) to select the required setting in selection field 1
(see table).
 Once the required setting has been selected, the Memory key (B)
lights up.
3. Save the setting.
4. Exit the service routine.
11 Commissioning, demo mode and repacking
11.12 Service routines for commissioning and repacking
Code
Function
00
Occlusal bite block function
deactivated*
01
Occlusal bite block function activated
147