Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
Selecting an X-ray component
64 95 142 D3632
D3632.031.01.02.02 03.2016
Automatic search
✔
The menu for adding X-ray components opens.
"Automatic search"
1. Click the
The network is searched for available X-ray components.
The X-ray components found on the network are displayed in list
form.
NOTE: If no X-ray component is found, check the availability of
the unit and repeat the
device manually"
2. Select the preferred X-ray component and enter
box in the location of the X-ray component.
"Save"
3. Click the
button to save the settings.
4. Wait until the saving operation has been completed.
5. Then click the button marked
The menu closes.
11 Commissioning, demo mode and repacking
11.8 Setting up the X-ray component
button in the structure tree.
"Automatic search"
function.
"Location"
"Exit"
.
"Add
or use the
in the text
115