13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
252
3. Overwrite the adjustment value for S1 in the text box of the
"Adjustment of Ceph secondary diaphragm (Quickshot)"
the measured value.
For information about the shifting direction (input of +/- sign in the
menu), see the chapter entitled "Shifting direction of the exposed
image area [ → 207]".
NOTICE! Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter
entitled "Automatic adjustment: Ceph secondary diaphragm
(QuickShot) (standard) [ → 247]"
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
menu with
64 95 142 D3632
D3632.031.01.02.02 03.2016