13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
Lx
Ly
D
246
3. Set the diaphragm tilt via screw (C) (Lx in mm) and the diaphragm
height via screw (D) (Ly in mm).
Screw (C):
Turn counterclockwise = adjusts the diaphragm to the right
Turn clockwise = adjusts the diaphragm to the left
Screw (D):
Turn counterclockwise = adjusts the diaphragm downward
Turn clockwise = adjusts the diaphragm upward
To measure the adjustment, use a gauge.
The above-mentioned directions of rotation for the screws apply to
both left-arm and right-arm versions.
4. Retighten screws (B) firmly.
5. Once adjustment is complete, put the secondary diaphragm cover
C
back again and fix it in place with the screw (A).
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016