13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
254
3. Create an exposure (80 kV / 14 mA; 14.9 s) [ → 209].
 Adjustment value S1 is calculated automatically from the
exposure and entered in the text box of the
main beam direction"
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
"Adjustment of Ceph
menu.
D3632.031.01.02.02 03.2016