11 Commissioning, demo mode and repacking
11.11 Using demo mode – operation without radiation release
126
11.11.2

Switching off demo mode

1. Switch off the unit.
DANGER
Perilous shock hazard!
It is essential to switch off the unit and to wait at least another 4 minutes
before taking off the covers of the X-ray tube assembly.
2. Remove the "Rear tube assembly" cover.
3. Remove the cover plate of board DX6.
4. Set the dip switch S2 (DX6) to position 1.
5. Connect cable L5 (XRAY) to connector J6/J103 (DX6).
 Radiation release is now once again possible.
6. Switch on the unit and check the mode on the info screen.
Demo mode: ON means that: Demo mode is switched on (radiation
release is not possible)
Demo mode: OFF means that: Demo mode is switched off
(radiography, X-ray radiation are possible!)
7. Switch the unit off again and reattach the cover plate and the tube
assembly cover by following the dismantling procedure in reverse
order.