13.2 Adjustment and calibration via the calibration menu
A
Adjustment OK
B
Adjustment not OK
4. Repeat the procedure starting at Step 2 until the value in the menu's
text box is equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT:If automatic adjustment fails to produce values equal to
zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 242].
5. If adjustment is OK (the adjustment values in the text boxes are equal
to "zero"), save the values. [ → 209]
 The status indicator that prefixes the
secondary diaphragm (Quickshot)"
6. Continue with the next stage of the adjustment procedure. [ → 253]
13 Unit adjustment and calibration
● The exposed image section must lie
centered and straight in the image
section as well as inside the
superimposed auxiliary lines.
● A surrounding white border must be
present.
● The maximum density must lie in the
center of the diaphragm area.
"Adjustment of Ceph
element is checked.
249