12 Startup, measurements and tests (for USA/Canada only)
12.7 Checking the tube current
Performing measurements
Analyzing measurements
186
4. DANGER! After switching off the unit, wait at least 4 minutes (LED
V500 on the DX6 must no longer be on) before removing jumper H
from the DX6 board.
Remove the jumper (H) from connector X302 on the DX6 board.
5. DANGER! Only use fully insulated measuring wires.
Connect the digital multimeter with the measuring wires to test points
MA- (X5-) and MA+ (X4+) at connector X302 on the DX6 board.
6. On the multimeter, select the current measuring range 20mA DC.
7. NOTICE! If the lid of the tube assembly cover is not attached, the ring
circulation is impeded and the unit can be damaged.
Temporarily install the lid of the tube assembly cover on the unit using
the screw.
8. DANGER! Do not touch any live components!
Set the main switch (A) to I (see also Operating Instructions).
9. Wait for approx. 1 minute.
10. Press the R key.
 The unit moves to its starting position.
1. Select the P1 program (see Operating Instructions).
2. Set the highest kV/mA level, 66kV/8mA (see Operating Instructions).
3. Make ready for exposure in SIDEXIS 4 (see Operating Instructions).
4. CAUTION! Activating the release button triggers X-rays.
Start the exposure by pressing the release button. Hold down the
release button until image acquisition is completed and the acoustic
signal that indicates the end of the exposure can be heard.
IMPORTANT
1 mA corresponds to a tube current of 1 mA. The permissible tolerance
is ± 20 %.
➢ Read the voltage value on the display of the multimeter.
 The tube current must be 8mA ± 1.6mA.
 If the measured value does
tolerance, replace the
 If the measured value falls within the permissible tolerance,
conclude the measurement.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
not
fall within the permissible
tube assembly
(see Service Manual).
D3632.031.01.02.02 03.2016