Adjustment Of The Ceph Secondary Diaphragm - Sirona ORTHOPHOS SL 2D Installation Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
238

Adjustment of the ceph secondary diaphragm

13.2.1.6
IMPORTANT
If you cannot correctly adjust the ceph secondary diaphragm via the
calibration menu, the Ceph secondary diagram must first be readjusted
mechanically [ → 245].
After mechanical readjustment has been completed, repeat adjustment
via the calibration menu.
13.2.1.6.1
Automatic adjustment: Ceph secondary diaphragm (standard)
NOTICE
Phantom for Ceph exposure
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the unit before a ceph exposure is taken; otherwise, the
sensor could collide with the phantom.
The pan needle phantom is
unit.
The ceph test phantom does
unit.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
not
in the pan bite block holder on the
not
connect to the cover of the sensor
D3632.031.01.02.02 03.2016